Electrochemical concentration of no-carrier-added [(18)F]fluoride from [(18)O]water in a disposable microfluidic cell for radiosynthesis of (18)F-labeled radiopharmaceuticals.
Saiki, Hidekazu; Iwata, Ren; Nakanishi, Hiroaki; Wong, Rebecca; Ishikawa, Yoichi; Furumoto, Shozo; Yamahara, Ryo; Sakamoto, Katsumasa; Ozeki, Eiichi.
Afiliação
  • Saiki H; Technology Research Laboratory, Shimadzu Corporation, Kyoto, Japan.
Appl Radiat Isot ; 68(9): 1703-8, 2010 Sep.
Article em En | MEDLINE | ID: mdl-20189817
ABSTRACT
The realization of the electrochemical method for microfluidic radiosynthesis is described for concentrating aqueous no-carrier-added [(18)F]fluoride into an aprotic solvent in a disposable microfluidic cell. Flowing aqueous [(18)F]fluoride was introduced into a disposable microfluidic cell (16microL) under an electric potential (10V), followed by anhydrous MeCN. The trapped [(18)F]fluoride was released in MeCN containing Kryptofix 222-KHCO(3) (ca. 60microL) under heat and a reversed potential (-2.5V). An automated module provided the [(18)F]fluoride ready for subsequent microfluidic radiosynthesis in overall radiochemical yields of 60% within 6min.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioisótopos de Flúor / Compostos Radiofarmacêuticos / Técnicas Analíticas Microfluídicas / Equipamentos Descartáveis / Eletroquímica / Marcação por Isótopo Idioma: En Revista: Appl Radiat Isot Ano de publicação: 2010 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioisótopos de Flúor / Compostos Radiofarmacêuticos / Técnicas Analíticas Microfluídicas / Equipamentos Descartáveis / Eletroquímica / Marcação por Isótopo Idioma: En Revista: Appl Radiat Isot Ano de publicação: 2010 Tipo de documento: Article