Evaluation of a simple and low cost potentiometric biosensor for pharmaceutical and in vivo adrenaline determination.
Biosens Bioelectron
; 26(2): 798-802, 2010 Oct 15.
Article
em En
| MEDLINE
| ID: mdl-20643538
In this work the polyphenol oxidase (PPO) was the main component of a biosensor for adrenaline determination. The activity of this enzyme was measured in several vegetables. Banana (Musa sp.) extracts presented better results with 974 UA (units of activity). The biosensor was constructed with a polyethylene tube (0.8 mm i.d.) filled with: carbon paste containing 50 UA of the PPO in phosphate buffer (pH=7.00) solution and vaseline as agglutinant. When the biosensor was applied in medicine samples it provided a linear range from 8.00×10(-9) to 8.00×10(-4) mol L(-1); the results obtained with the proposed method and the Brazilian Pharmacopoeia method were in agreement (t-test). When it was applied in blood samples, the matrix-matching calibration was used, and the linear range was from 8.00×10(-7) to 8.00×10(-3) mol L(-1). In vivo studies were also done. The obtained results for those electrodes, which were inserted in the jugular vein of Wistar rats, were very promising.
Texto completo:
1
Coleções:
01-internacional
Contexto em Saúde:
1_ASSA2030
Base de dados:
MEDLINE
Assunto principal:
Técnicas Biossensoriais
/
Epinefrina
/
Condutometria
Tipo de estudo:
Evaluation_studies
/
Health_economic_evaluation
Limite:
Animals
Idioma:
En
Revista:
Biosens Bioelectron
Ano de publicação:
2010
Tipo de documento:
Article