Air kerma to dose equivalent conversion coefficients not included in ISO 4037-3.
Behrens, Rolf.
Afiliação
  • Behrens R; Physikalisch-Technische Bundesanstalt, Bundesallee 100, Braunschweig, Germany. rolf.behrens@ptb.de
Radiat Prot Dosimetry ; 147(3): 373-9, 2011 Nov.
Article em En | MEDLINE | ID: mdl-21156781
ABSTRACT
The International Organization for Standardization (ISO) has issued a standard series on photon reference radiation qualities (ISO 4037). In this work, conversion coefficients from air kerma to dose equivalent quantities not included in ISO 4037-3 are supplied for the following quantities For H(p)(0.07) for X and gamma radiation qualities for the rod, the pillar and the slab phantom and for H(p)(3) for X and gamma radiation qualities for the slab phantom. In addition, an overview of conversion coefficients suggested for use for all quantities relevant in radiation protection is provided.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Eficiência Biológica Relativa / Monitoramento de Radiação / Fótons / Imagens de Fantasmas / Ar Limite: Humans Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2011 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Eficiência Biológica Relativa / Monitoramento de Radiação / Fótons / Imagens de Fantasmas / Ar Limite: Humans Idioma: En Revista: Radiat Prot Dosimetry Ano de publicação: 2011 Tipo de documento: Article