Sinking and fit of abutment of locking taper implant system.
Moon, Seung-Jin; Kim, Hee-Jung; Son, Mee-Kyoung; Chung, Chae-Heon.
Afiliação
  • Moon SJ; Graduate student, Department of Prosthodontics, College of Dentistry, Chosun University, Gwangju, Korea.
J Adv Prosthodont ; 1(2): 97-101, 2009 Jul.
Article em En | MEDLINE | ID: mdl-21165262
ABSTRACT
STATEMENT OF

PROBLEM:

Unlike screw-retention type, fixture-abutment retention in Locking taper connection depends on frictional force so it has possibility of abutment to sink.

PURPOSE:

In this study, Bicon® Implant System, one of the conical internal connection implant system, was used with applying loading force to the abutments connected to the fixture. Then the amount of sinking was measured. MATERIAL AND

METHODS:

10 Bicon® implant fixtures were used. First, the abutment was connected to the fixture with finger force. Then it was tapped with a mallet for 3 times and loads of 20 kg corresponding to masticatory force using loading application instrument were applied successively. The abutment state, slightly connected to the fixture without pressure was considered as a reference length, and every new abutment length was measured after each load's step was added. The amount of abutment sinking (mm) was gained by subtracting the length of abutment-fixture under each loading condition from reference length.

RESULTS:

It was evident, that the amount of abutment sinking in Bicon® Implant System increased as loads were added. When loads of 20 kg were applied more than 5 - 7 times, sinking stopped at 0.45 ± 0.09 mm.

CONCLUSION:

Even though locking taper connection type implant shows good adaption to occlusal force, it has potential for abutment sinking as loads are given. When locking taper connection type implant is used, satisfactory loads are recommended for precise abutment location.