Empirical formula for the prediction of off axis ratios and isodose curves for a treatment planning system.
J Med Phys
; 31(4): 262-8, 2006 Oct.
Article
em En
| MEDLINE
| ID: mdl-21206642
A mathematical model has been developed for prediction of off axis ratio (OAR), using Wood - Saxon term used to represent nuclear potential. This method has been satisfactorily applied for predicting OAR in case of (60)Co γ-rays and high energy X-rays. Investigations are considered upto a depth of 25 cm in the case of 4MV LINAC for which measurements were carried out in our laboratory using indigenously developed Radiation Field Analyzer. For (60)Co γ-rays as well as 6 and 18MV LINAC beams we could get off-axis profiles only upto 20 cm. The shift δ between measured and predicted OAR is within ±2 mm except for 20 cm depth near the falling edge of the penumbra, where it is 2.80 mm. Software has been developed in Visual Basic 6 on Windows platform to plot Isodose curves, which is based on the mathematical modeling of OAR and central axis percentage depth dose.