Your browser doesn't support javascript.
loading
Empirical formula for the prediction of off axis ratios and isodose curves for a treatment planning system.
Pal, Surajit; Ravishankar, R; Sharma, R P; Muthukrishnan, G; Ray, Dilip Kr; Roy, S N; Srivastava, D K.
Afiliação
  • Pal S; Variable Energy Cyclotron Centre, 1/AF Bidhan Nagar, Kolkata, India.
J Med Phys ; 31(4): 262-8, 2006 Oct.
Article em En | MEDLINE | ID: mdl-21206642
A mathematical model has been developed for prediction of off axis ratio (OAR), using Wood - Saxon term used to represent nuclear potential. This method has been satisfactorily applied for predicting OAR in case of (60)Co γ-rays and high energy X-rays. Investigations are considered upto a depth of 25 cm in the case of 4MV LINAC for which measurements were carried out in our laboratory using indigenously developed Radiation Field Analyzer. For (60)Co γ-rays as well as 6 and 18MV LINAC beams we could get off-axis profiles only upto 20 cm. The shift δ between measured and predicted OAR is within ±2 mm except for 20 cm depth near the falling edge of the penumbra, where it is 2.80 mm. Software has been developed in Visual Basic 6 on Windows platform to plot Isodose curves, which is based on the mathematical modeling of OAR and central axis percentage depth dose.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Prognostic_studies / Risk_factors_studies Idioma: En Revista: J Med Phys Ano de publicação: 2006 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Prognostic_studies / Risk_factors_studies Idioma: En Revista: J Med Phys Ano de publicação: 2006 Tipo de documento: Article