A whole-body dosimetry system for personal monitoring based on hot-pressed thin layer TLD.
Radiat Prot Dosimetry
; 144(1-4): 246-8, 2011 Mar.
Article
em En
| MEDLINE
| ID: mdl-21227958
We are introducing a new high-capacity thermoluminescent dosemeter (TLD) system to measure the whole body values of H(p)(10) and H(p)(0.07) from photons for use in individual monitoring services. Small and light-weight badges allow a convenient application in a wide variety of workplaces with photon radiation from 20 keV to at least 7 MeV. The main advantage of this system will be the large capacity of â¼ 100,000 dosemeters per month at costs equivalent to the current film monitoring. The hot-pressed thin-layer TL detector (LiF:Mg,Ti) is welded onto an aluminium substrate and provided with a data matrix code for automatic processing. The detector holder has been optimised, that no additional filter is necessary. The new designed TLD reader with readout times <10 s will allow a large throughput and a considerable degree of automation.