Determination of octreotide and assessment of matrix effects in human plasma using ultra high performance liquid chromatography-tandem mass spectrometry.
J Chromatogr B Analyt Technol Biomed Life Sci
; 879(22): 2081-8, 2011 Jul 15.
Article
em En
| MEDLINE
| ID: mdl-21700513
A selective UHPLC-MS/MS method for determination of the therapeutic peptide octreotide in human plasma was developed and validated. This assay used a UHPLC C(18) column with 1.7 µm particle size for efficient separation and an ion-exchange SPE for selective extraction. Octreotide and its labeled internal standard, [(13)C(6)Phe(3)] octreotide, were extracted from human plasma using a simple Oasis® WCX µElution SPE method and analyzed with a total chromatographic run time of 7.5 min. Matrix effects were studied during method development by direct monitoring of representative phospholipids. On-line removal of phospholipids using column switching and pre-column back-flushing was carried out to trap and remove any residual phospholipid matrix interferences. The UHPLC column provided baseline separation between the analyte and matrix peaks. The chromatographic conditions yielded optimal retention and excellent peak shape for both the analyte and internal standard. The assay was linear in the concentration range of 0.025-25.0 ng/ml, inter- and intra-assay precision and accuracy were within 6.1% and ±1.93%, respectively. Recovery was â¼73%. Post-extraction addition experiments showed that matrix effects were less than 4%. This method for octreotide in human plasma has been validated and utilized to support of clinical pharmacokinetic studies.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Octreotida
/
Cromatografia Líquida de Alta Pressão
/
Espectrometria de Massas em Tandem
Tipo de estudo:
Diagnostic_studies
Limite:
Humans
Idioma:
En
Revista:
J Chromatogr B Analyt Technol Biomed Life Sci
Ano de publicação:
2011
Tipo de documento:
Article