Biomechanical behavior of a biomimetic artificial intervertebral disc.
Spine (Phila Pa 1976)
; 37(6): E367-73, 2012 Mar 15.
Article
em En
| MEDLINE
| ID: mdl-21912317
STUDY DESIGN: The biomechanical behavior of a biomimetic artificial intervertebral disc (AID) was characterized in vitro in axial compression and compared with natural disc behavior. OBJECTIVE: To evaluate the strength and durability of a novel biomimetic AID and to demonstrate whether its axial deformation behavior is similar to that of a natural disc. SUMMARY OF BACKGROUND DATA: Current clinically used AIDs have reasonable success rates. However, because of their nonphysiological design, spinal mechanics are altered. To avoid long-term complications, a novel biomimetic AID, with a nucleus-annulus structure and osmotic swelling properties has been developed. METHODS: Eighteen AIDs in total were tested in axial compression. Six were loaded monotonically to determine strength. Six were tested in fatigue (600-6000 N). Another 6 were used to characterize the axial creep and dynamic behavior (0.01-10 Hz). Creep and dynamic response were also determined for 4 AIDs after fatigue loading. RESULTS: The AIDs remained intact up to 15 kN and 10 million cycles. The creep and dynamic behavior were similar to the natural disc behavior, except for hysteresis, which was 20% to 30% higher. After fatigue, creep decreased from 4% to 1%, stiffness increased 2-fold, and hysteresis was reduced to that for a normal disc. CONCLUSION: A strong and durable AID design was introduced. Compared with current clinical articulating AIDs, this biomimetic AID introduces the natural disc annulus-nucleus structure, resulting in axial behavior closer to that of the natural disc.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Próteses e Implantes
/
Fenômenos Biomecânicos
/
Força Compressiva
/
Disco Intervertebral
Limite:
Humans
Idioma:
En
Revista:
Spine (Phila Pa 1976)
Ano de publicação:
2012
Tipo de documento:
Article