Your browser doesn't support javascript.
loading
Determining the risks of magnetic resonance imaging at 1.5 tesla for patients with pacemakers and implantable cardioverter defibrillators.
Cohen, Jennifer D; Costa, Heather S; Russo, Robert J.
Afiliação
  • Cohen JD; Division of Cardiovascular Diseases, Scripps Clinic, La Jolla, California, USA.
Am J Cardiol ; 110(11): 1631-6, 2012 Dec 01.
Article em En | MEDLINE | ID: mdl-22921995
ABSTRACT
Conventional pacemaker and implantable cardioverter-defibrillator product labeling currently cautions against exposure to magnetic resonance imaging (MRI). However, there is a growing clinical need for MRI, without an acceptable alternative imaging modality in many patients with cardiac devices. The purpose of this study was to determine the risk of MRI at 1.5 T for patients with cardiac devices by measuring the frequency of device failures and clinically relevant device parameter changes. Data from a single-center retrospective review of 109 patients with pacemakers and implantable cardioverter-defibrillators (the MRI group) who underwent 125 clinically indicated MRI studies were compared to data from a prospective cohort of 50 patients with cardiac devices who did not undergo MRI (the control group). In the MRI group, there were no deaths, device failures requiring generator or lead replacement, induced arrhythmias, losses of capture, or electrical reset episodes. Decreases in battery voltage of ≥0.04 V occurred in 4%, pacing threshold increases of ≥0.5 V in 3%, and pacing lead impedance changes of ≥50 Ω in 6%. Although there were statistically significant differences between the MRI and control groups for the mean change in pacing lead impedance (-6.2 ± 23.9 vs 3.0 ± 22.1 Ω) and left ventricular pacing threshold (-0.1 ± 0.3 vs 0.1 ± 0.2 V), these differences were not clinically important. In conclusion, MRI in patients with cardiac devices resulted in no device or lead failures. A small number of clinically relevant changes in device parameter measurements were noted. However, these changes were similar to those in a control group of patients who did not undergo MRI.
Assuntos

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 6_ODS3_enfermedades_notrasmisibles Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis / Medição de Risco / Imagem Cinética por Ressonância Magnética / Falha de Equipamento Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Am J Cardiol Ano de publicação: 2012 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 6_ODS3_enfermedades_notrasmisibles Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Arritmias Cardíacas / Desfibriladores Implantáveis / Medição de Risco / Imagem Cinética por Ressonância Magnética / Falha de Equipamento Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: Am J Cardiol Ano de publicação: 2012 Tipo de documento: Article