Design and validation of an indicator gas injector for multiple gas washout tests in mechanically ventilated patients.
Huygen, P E; Feenstra, B W; Holland, W P; Ince, C; Stam, H; Bruining, H A.
Afiliação
  • Huygen PE; Department of Surgery, University Hospital Dijkzigt, Rotterdam, The Netherlands.
Crit Care Med ; 18(7): 754-9, 1990 Jul.
Article em En | MEDLINE | ID: mdl-2364717
ABSTRACT
A device to produce a stepwise indicator gas fraction variation to initiate a washout test in mechanically ventilated patients is described. The device, which can be used in conjunction with the commonly used Siemens-Elema series 900 ventilators, is based on simple, off-the-shelf technology. It features the simultaneous use of two indicator gases (so that the influence of diffusion processes in the gas exchange to the patient can be measured) and maintains a nearly constant FIO2 during a washout procedure. With this indicator gas injector, the transition time of the indicator gas fraction at the beginning of a washout proved to be short enough to detect ventilation inhomogeneity by visual inspection of the washout curves. Functional residual capacity measurements using this device are presented on a test lung with known volume, on healthy volunteers, and on critically ill patients.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração / Ventiladores Mecânicos Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Crit Care Med Ano de publicação: 1990 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração / Ventiladores Mecânicos Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Crit Care Med Ano de publicação: 1990 Tipo de documento: Article