Your browser doesn't support javascript.
loading
In vitro biomechanical testing of a micro external skeletal fixator.
Deiss, R; Bali, M S; Doherr, M; Spreng, D; Rytz, U; Ferguson, S J.
Afiliação
  • Deiss R; Dr. med. vet. Ramona Deiss, Bahnhofstrasse 56, 3127 Mühlethurnen, Switzerland, Phone: +41 79 786 86 32, E-mail: ramona.deiss@me.com.
Vet Comp Orthop Traumatol ; 26(5): 385-91, 2013.
Article em En | MEDLINE | ID: mdl-23708943
ABSTRACT

OBJECTIVE:

To biomechanically test the properties of three different Universal Micro External Fixator (UMEX™) configurations with regard to their use in very small animals (<5kg) and compare the UMEX system to the widely used IMEX External Skeletal Fixation (SK™) system in terms of stiffness, space needed for pin placement and weight.

METHODS:

Three different UMEX configurations (type Ia, type Ib, and type II modified) and one SK configuration type Ia were used to stabilize Delrin plastic rods in a 1 cm fracture gap model. These constructs were tested in axial compression, craniocaudal bending, mediolateral bending, and torsion. Testing was conducted within the elastic range and mean stiffness in each mode was determined from the slope of the linear portion of the load-deformation curve. A Kruskal Wallis one-way analysis of variance on ranks test was utilized to assess differences between constructs (p <0.05).

RESULTS:

The UMEX type II modified configuration was significantly stiffer than the other UMEX configurations and the SK type Ia, except in craniocaudal bending, where the SK type Ia configuration was stiffer than all UMEX constructs. The UMEX type Ia configuration was significantly the weakest of those frames. The UMEX constructs were lighter and smaller than the SK, thus facilitating closer pin placement.

CONCLUSIONS:

Results supported previous reports concerning the superiority of more complex constructs regarding stiffness. The UMEX system appears to be a valid alternative for the treatment of long-bone fractures in very small animals.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Teste de Materiais / Fixadores Externos / Falha de Equipamento Limite: Animals Idioma: En Revista: Vet Comp Orthop Traumatol Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Teste de Materiais / Fixadores Externos / Falha de Equipamento Limite: Animals Idioma: En Revista: Vet Comp Orthop Traumatol Ano de publicação: 2013 Tipo de documento: Article