Your browser doesn't support javascript.
loading
Assessment of the performances of AcuStar HIT and the combination with heparin-induced multiple electrode aggregometry: a retrospective study.
Minet, V; Bailly, N; Douxfils, J; Osselaer, J C; Laloy, J; Chatelain, C; Elalamy, I; Chatelain, B; Dogné, J M; Mullier, F.
Afiliação
  • Minet V; Department of Pharmacy, Namur Research Institute for LIfe Sciences (NARILIS), Namur Thrombosis and Hemostasis Center (NTHC), University of Namur, Namur, Belgium. Electronic address: valentine.minet@unamur.be.
Thromb Res ; 132(3): 352-9, 2013 Sep.
Article em En | MEDLINE | ID: mdl-23803389
BACKGROUND: Early diagnosis of immune heparin-induced thrombocytopenia (HIT) is challenging. HemosIL® AcuStar HIT and heparin-induced multiple electrode aggregometry (HIMEA) were recently proposed as rapid diagnostic methods. OBJECTIVES: We conducted a study to assess performances of AcuStar HIT-IgG (PF4-H) and AcuStar HIT-Ab (PF4-H). The secondary objective was to compare the performances of the combination of Acustar HIT and HIMEA with standardised clinical diagnosis. METHODS: Sera of 104 suspected HIT patients were retrospectively tested with AcuStar HIT. HIMEA was performed on available sera (n=81). The clinical diagnosis was established by analysing in a standardized manner the patient's medical records. These tests were also compared with PF4-Enhanced®, LTA, and SRA in subsets of patients. Thresholds were determined using ROC curve analysis with clinical outcome as reference. RESULTS: Using the recommended thresholds (1.00AU), the negative predictive value (NPV) of HIT-IgG and HIT-Ab were 100.0% (95% CI: 95.9%-100.0% and 95.7%-100.0%). The positive predictive value (PPV) were 64.3% (95% CI: 35.1%-87.2.2%) and 45.0% (95% CI: 23.2%-68.6%), respectively. Using our thresholds (HIT-IgG: 2.89AU, HIT-Ab: 9.41AU), NPV of HIT-IgG and HIT-Ab were 100.0% (95% CI: 96.0%-100.0% and 96.1%-100.0%). PPV were 75.0% (95% CI: 42.7%-94.5%) and 81.8% (95% CI: 48.3%-97.7%), respectively. Of the 79 patients with a medium-high pretest probability score, 67 were negative using HIT-IgG (PF4-H) test at our thresholds. HIMEA was performed on HIT-IgG positive patients. Using this combination, only one patient on 79 was incorrectly diagnosed. CONCLUSION: Acustar HIT showed good performances to exclude the diagnosis of HIT. Combination with HIMEA improves PPV.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Função Plaquetária / Trombocitopenia / Heparina / Agregação Plaquetária / Medições Luminescentes Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Thromb Res Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Função Plaquetária / Trombocitopenia / Heparina / Agregação Plaquetária / Medições Luminescentes Tipo de estudo: Observational_studies / Prognostic_studies / Risk_factors_studies / Screening_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Thromb Res Ano de publicação: 2013 Tipo de documento: Article