ABSTRACT
STUDY

OBJECTIVE:

The Manufacturer and User Facility Device Experience database may be useful for clinicians using a Food and Drug Administration-approved medical device to identify the occurrence of adverse events and complications. We sought to analyze and investigate reports associated with the Essure hysteroscopic sterilization system (Conceptus Inc., Mountain View, CA) using this database.

DESIGN:

Retrospective review of the Manufacturer and User Facility Device Experience database for events related to Essure hysteroscopic sterilization from November 2002 to February 2012 (Canadian Task Force Classification III).

SETTING:

Online retrospective review. PATIENTS Online reports of patients who underwent Essure tubal sterilization. INTERVENTION Essure tubal sterilization. MEASUREMENTS AND MAIN

RESULTS:

Four hundred fifty-seven adverse events were reported in the study period. Pain was the most frequently reported event (217 events [47.5%]) followed by delivery catheter malfunction (121 events [26.4%]). Poststerilization pregnancy was reported in 61 events (13.3%), of which 29 were ectopic pregnancies. Other reported events included perforation (90 events [19.7%]), abnormal bleeding (44 events [9.6%]), and microinsert malposition (33 events [7.2%]). The evaluation and management of these events resulted in an additional surgical procedure in 270 cases (59.1%), of which 44 were hysterectomies.

CONCLUSION:

Sixty-one unintended poststerilization pregnancies were reported in the study period, of which 29 (47.5%) were ectopic gestations. Thus, ectopic pregnancy must be considered if a woman becomes pregnant after Essure hysteroscopic sterilization. Additionally, 44 women underwent hysterectomy after an adverse event reported to be associated with the use of the device.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dor / Gravidez Ectópica / Esterilização Reprodutiva / Histeroscopia Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Adult / Female / Humans / Pregnancy Idioma: En Revista: J Minim Invasive Gynecol Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Dor / Gravidez Ectópica / Esterilização Reprodutiva / Histeroscopia Tipo de estudo: Observational_studies / Risk_factors_studies Limite: Adult / Female / Humans / Pregnancy Idioma: En Revista: J Minim Invasive Gynecol Ano de publicação: 2013 Tipo de documento: Article