Your browser doesn't support javascript.
loading
Feasibility of a portable bihormonal closed-loop system to control glucose excursions at home under free-living conditions for 48 hours.
van Bon, Arianne C; Luijf, Yoeri M; Koebrugge, Rob; Koops, Robin; Hoekstra, Joost B L; DeVries, J Hans.
Afiliação
  • van Bon AC; 1 Department of Internal Medicine, Academic Medical Centre , Amsterdam, The Netherlands .
Diabetes Technol Ther ; 16(3): 131-6, 2014 Mar.
Article em En | MEDLINE | ID: mdl-24224750
BACKGROUND: This study assessed the feasibility of a portable bihormonal closed-loop system at home. SUBJECTS AND METHODS: Sixteen pump-treated patients with type 1 diabetes received 48 h of closed-loop therapy with a telemonitored insulin- and glucagon-delivering closed-loop system and 48 h of patient-managed open-loop therapy. RESULTS: Owing to technical problems in five cases, only 11 patients could be analyzed. Whereas median (interquartile range) glucose levels were not significantly different during Day 1 of open-loop control (OL1) from closed-loop control (CL1) (8.27 [0.83] mmol/L vs. 8.84 [1.47] mmol/L; P=0.206), they were significantly lower during Day 2 of closed-loop control (CL2) versus open-loop control (OL2) (7.70 [2.29] mmol/L vs. 8.84 [0.87] mmol/L; P=0.027). Time spent in euglycemia (3.9-10 mmol/L) was comparable with 67.2% (38.5%) in OL1 versus 79.2% (16.9%) in CL1 (P=0.189) and 66.0% (29.8%) in OL2 versus 76.5% (23.9%) in CL2 (P=0.162). Time spent in hypoglycemia (<3.9 mmol/L) was comparable on Day 1 of control (OL1, 0.68% [8.68%]; CL1, 2.08% [7.61%]; P=0.593) but significantly higher during Day 2 of control (OL2, 0.00% [11.07%]; CL2, 2.8% [9.8%]; P=0.0172) (P=0.017). CONCLUSIONS: Bihormonal closed-loop control is feasible at home, with comparable time in euglycemia to open-loop control and significantly lower median glucose levels on Day 2 of control at the expense of more time in hypoglycemia, albeit still at a very low percentage of time.
Assuntos

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 / 2_ODS3 Base de dados: MEDLINE Assunto principal: Glicemia / Hemoglobinas Glicadas / Sistemas de Infusão de Insulina / Monitorização Ambulatorial / Diabetes Mellitus Tipo 1 / Hipoglicemiantes / Insulina Tipo de estudo: Evaluation_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Diabetes Technol Ther Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 / 2_ODS3 Base de dados: MEDLINE Assunto principal: Glicemia / Hemoglobinas Glicadas / Sistemas de Infusão de Insulina / Monitorização Ambulatorial / Diabetes Mellitus Tipo 1 / Hipoglicemiantes / Insulina Tipo de estudo: Evaluation_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Diabetes Technol Ther Ano de publicação: 2014 Tipo de documento: Article