Your browser doesn't support javascript.
loading
Isotope-labelled urea to test colon drug delivery devices in vivo: principles, calculations and interpretations.
Maurer, Marina J M; Schellekens, Reinout C A; Wutzke, Klaus D; Stellaard, Frans.
Afiliação
  • Maurer MJ; a Department of Clinical and Hospital Pharmacy , University Medical Center Groningen , Groningen , The Netherlands.
Isotopes Environ Health Stud ; 49(4): 473-91, 2013.
Article em En | MEDLINE | ID: mdl-24313370
ABSTRACT
This paper describes various methodological aspects that were encountered during the development of a system to monitor the in vivo behaviour of a newly developed colon delivery device that enables oral drug treatment of inflammatory bowel diseases. [(13)C]urea was chosen as the marker substance. Release of [(13)C]urea in the ileocolonic region is proven by the exhalation of (13)CO2 in breath due to bacterial fermentation of [(13)C]urea. The (13)CO2 exhalation kinetics allows the calculation of a lag time as marker for delay of release, a pulse time as marker for the speed of drug release and the fraction of the dose that is fermented. To determine the total bioavailability, also the fraction of the dose absorbed from the intestine must be quantified. Initially, this was done by calculating the time-dependent [(13)C]urea appearance in the body urea pool via measurement of (13)C abundance and concentration of plasma urea. Thereafter, a new methodology was successfully developed to obtain the bioavailability data by measurement of the urinary excretion rate of [(13)C]urea. These techniques required two experimental days, one to test the coated device, another to test the uncoated device to obtain reference values for the situation that 100 % of [(13)C]urea is absorbed. This is hampered by large day-to-day variations in urea metabolism. Finally, a completely non-invasive, one-day test was worked out based on a dual isotope approach applying a simultaneous administration of [(13)C]urea in a coated device and [(15)N2]urea in an uncoated device. All aspects of isotope-related analytical methodologies and required calculation and correction systems are described.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ureia / Sistemas de Liberação de Medicamentos / Colo Limite: Humans Idioma: En Revista: Isotopes Environ Health Stud Ano de publicação: 2013 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ureia / Sistemas de Liberação de Medicamentos / Colo Limite: Humans Idioma: En Revista: Isotopes Environ Health Stud Ano de publicação: 2013 Tipo de documento: Article