Static field influences on transcranial magnetic stimulation: considerations for TMS in the scanner environment.
Yau, Jeffrey M; Jalinous, Reza; Cantarero, Gabriela L; Desmond, John E.
Afiliação
  • Yau JM; Department of Neurology, Johns Hopkins Medical Institutions, Baltimore, MD 21205, USA. Electronic address: jeffrey.yau@bcm.edu.
  • Jalinous R; The Magstim Company Ltd, Whitland, Wales, UK.
  • Cantarero GL; Solomon H. Snyder Department of Neuroscience, Johns Hopkins Medical Institutions, Baltimore, MD 21205, USA; Department of Physical Medicine and Rehabilitation, Johns Hopkins Medical Institutions, Baltimore, MD 21205, USA.
  • Desmond JE; Department of Neurology, Johns Hopkins Medical Institutions, Baltimore, MD 21205, USA; Solomon H. Snyder Department of Neuroscience, Johns Hopkins Medical Institutions, Baltimore, MD 21205, USA.
Brain Stimul ; 7(3): 388-93, 2014.
Article em En | MEDLINE | ID: mdl-24656916
ABSTRACT

BACKGROUND:

Transcranial magnetic stimulation (TMS) can be combined with functional magnetic resonance imaging (fMRI) to simultaneously manipulate and monitor human cortical responses. Although tremendous efforts have been directed at characterizing the impact of TMS on image acquisition, the influence of the scanner's static field on the TMS coil has received limited attention. OBJECTIVE/

HYPOTHESIS:

The aim of this study was to characterize the influence of the scanner's static field on TMS. We hypothesized that spatial variations in the static field could account for TMS field variations in the scanner environment.

METHODS:

Using an MRI-compatible TMS coil, we estimated TMS field strengths based on TMS-induced voltage changes measured in a search coil. We compared peak field strengths obtained with the TMS coil positioned at different locations (B0 field vs fringe field) and orientations in the static field. We also measured the scanner's static field to derive a field map to account for TMS field variations.

RESULTS:

TMS field strength scaled depending on coil location and orientation with respect to the static field. Larger TMS field variations were observed in fringe field regions near the gantry as compared to regions inside the bore or further removed from the bore. The scanner's static field also exhibited the greatest spatial variations in fringe field regions near the gantry.

CONCLUSIONS:

The scanner's static field influences TMS fields and spatial variations in the static field correlate with TMS field variations. Coil orientation changes in the B0 field did not result in substantial TMS field variations. TMS field variations can be minimized by delivering TMS in the bore or outside of the 0-70 cm region from the bore entrance.