Validation of the V(max) metabolic cart in a simulated pediatric model.
Wines, Kristen N; Rzepecki, Alexandra K; Andrews, Audrey L; Dechert, Ronald E.
Afiliação
  • Wines KN; C. S. Mott Children's Hospital & Von Voigtlander Women's Hospital, Ann Arbor, Michigan kwines@umich.edu.
  • Rzepecki AK; C. S. Mott Children's Hospital & Von Voigtlander Women's Hospital, Ann Arbor, Michigan.
  • Andrews AL; C. S. Mott Children's Hospital & Von Voigtlander Women's Hospital, Ann Arbor, Michigan.
  • Dechert RE; C. S. Mott Children's Hospital & Von Voigtlander Women's Hospital, Ann Arbor, Michigan.
JPEN J Parenter Enteral Nutr ; 39(3): 353-8, 2015 Mar.
Article em En | MEDLINE | ID: mdl-24658462
ABSTRACT

BACKGROUND:

The ability to accurately measure resting energy expenditure (REE) using indirect calorimetry, often referred to as the "gold standard" in nutrition needs assessment, is important given the well-established positive correlation between patient outcome and proportion of the nutrition goal met. While many studies have been done to compare various metabolic carts with one another, the literature lacks a large simulator-based validation of any metabolic cart system. MATERIALS AND

METHODS:

In the present study, 8 specifically trained staff members independently conducted 10 simulation trials each using the V(max) Encore metabolic analyzer in conjunction with a metabolic calibration system, which simulates patient metabolic activity, to validate the accuracy of the V(max) Encore across a wide range of simulated metabolic conditions. Testing conditions consisted of incremental adjustments in calibrated gas infusion with a consistently set tidal volume and respiratory rate.

RESULTS:

There was a strong, statistically significant correlation between the predicted and actual VO2 and VCO2 data (VO2, R (2) = 0.998; VCO2, R (2) = 0.997). In addition, we observed no significant difference between individuals performing these trials (VO2, P = 1.000, F = 0.021, df = 79; VCO2, P = 1.000, F = 0.030, df = 79).

CONCLUSIONS:

This study is the first to report on such a wide spectrum of metabolic activity (50-2000 kcal REE) using a calibrated bench model and validates the accuracy, reproducibility, and use of the V(max) Encore metabolic cart.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Consumo de Oxigênio / Descanso / Metabolismo Basal / Calorimetria Indireta Tipo de estudo: Prognostic_studies Limite: Child / Humans Idioma: En Revista: JPEN J Parenter Enteral Nutr Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Consumo de Oxigênio / Descanso / Metabolismo Basal / Calorimetria Indireta Tipo de estudo: Prognostic_studies Limite: Child / Humans Idioma: En Revista: JPEN J Parenter Enteral Nutr Ano de publicação: 2015 Tipo de documento: Article