Bernert, M; Eich, T; Burckhart, A; Fuchs, J C; Giannone, L; Kallenbach, A; McDermott, R M; Sieglin, B.
Afiliação
  • Bernert M; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • Eich T; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • Burckhart A; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • Fuchs JC; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • Giannone L; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • Kallenbach A; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • McDermott RM; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
  • Sieglin B; Max Planck Institute for Plasma Physics, Boltzmannstr. 2, 85748 Garching, Germany.
Rev Sci Instrum ; 85(3): 033503, 2014 Mar.
Article em En | MEDLINE | ID: mdl-24689581
ABSTRACT
In the ASDEX Upgrade tokamak, a radiation measurement for a wide spectral range, based on semiconductor detectors, with 256 lines of sight and a time resolution of 5 µs was recently installed. In combination with the foil based bolometry, it is now possible to estimate the absolutely calibrated radiated power of the plasma on fast timescales. This work introduces this diagnostic based on AXUV (Absolute eXtended UltraViolet) n-on-p diodes made by International Radiation Detectors, Inc. The measurement and the degradation of the diodes in a tokamak environment is shown. Even though the AXUV diodes are developed to have a constant sensitivity for all photon energies (1 eV-8 keV), degradation leads to a photon energy dependence of the sensitivity. The foil bolometry, which is restricted to a time resolution of less than 1 kHz, offers a basis for a time dependent calibration of the diodes. The measurements of the quasi-calibrated diodes are compared with the foil bolometry and found to be accurate on the kHz time scale. Therefore, it is assumed, that the corrected values are also valid for the highest time resolution (200 kHz). With this improved diagnostic setup, the radiation induced by edge localized modes is analyzed on fast timescales.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2014 Tipo de documento: Article