Your browser doesn't support javascript.
loading
Improved durability of flex-Xc digital flexible ureteroscope: how long can you expect it to last?
Multescu, Razvan; Geavlete, Bogdan; Georgescu, Dragos; Geavlete, Petrisor.
Afiliação
  • Multescu R; Department of Urology, Saint John Clinical Emergency Hospital, Bucharest, Romania. Electronic address: razvanmultescu@yahoo.com.
  • Geavlete B; Department of Urology, Saint John Clinical Emergency Hospital, Bucharest, Romania.
  • Georgescu D; Department of Urology, Saint John Clinical Emergency Hospital, Bucharest, Romania.
  • Geavlete P; Department of Urology, Saint John Clinical Emergency Hospital, Bucharest, Romania.
Urology ; 84(1): 32-5, 2014 Jul.
Article em En | MEDLINE | ID: mdl-24703462
ABSTRACT

OBJECTIVE:

To retrospectively analyze a significant number of flexible ureterorenoscopic procedures performed with the Storz Flex-Xc model to evaluate its durability. MATERIALS AND

METHODS:

Between May 2012 and May 2013, all the flexible ureteroscopic procedures performed with 3 Storz Flex-Xc flexible ureteroscopes were analyzed. Relocation of the lower pole stones was performed in all cases when possible. In the second and third series, the access sheath was retracted together with the flexible ureteroscope, thus preventing the damages, which its distal part may inflict on the flexed endoscope.

RESULTS:

A total of 406 procedures were performed on 372 patients first endoscope used on 96 procedures (90 patients), second one on 151 procedures (139 patients), and third one on 159 procedures (143 patients). Ureteral access sheath was used in 71.4% of the cases. The first endoscope was used for 67.1 hours, the second for 107.7 hours, and the third one for 107.2 hours. Major repairs were needed after significant damages of the outer coating (first endoscope) and severe deterioration of the deflecting mechanism (second and third endoscopes).

CONCLUSION:

The digital Storz Flex-Xc seems to be a durable model of flexible ureteroscope. The use of ureteral access sheath and avoidance of overstressing the deflection mechanism by relocating lower pole stones seem to offer substantial advantages in prolonging the lifespan of these expensive instruments.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cálculos Ureterais / Ureteroscópios Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Urology Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cálculos Ureterais / Ureteroscópios Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Urology Ano de publicação: 2014 Tipo de documento: Article