Comparison of strength, consistency, and speed of COR-KNOT versus manually hand-tied knots in an ex vivo minimally invasive model.
Lee, Candice Y; Sauer, Jude S; Gorea, Heather R; Martellaro, Angelo J; Knight, Peter A.
Afiliação
  • Lee CY; From the *University of Rochester Medical Center, Rochester, NY USA; and †LSI SOLUTIONS, Victor, NY USA.
Innovations (Phila) ; 9(2): 111-6; discussion 116, 2014.
Article em En | MEDLINE | ID: mdl-24755535
ABSTRACT

OBJECTIVE:

This study compared the strength, consistency, and speed of prosthetic attachment sutures secured with automated fasteners with those of manual knots using an ex vivo porcine mitral valve annuloplasty model. A novel miniature pressure transducer system was developed to quantify pressures between sutured prosthetic rings and underlying cardiac tissue.

METHODS:

Sixteen mitral annuloplasty rings were sewn into ex vivo pig hearts. Eight rings were secured with the COR-KNOT device; and eight rings, with hand-tied knots using a knot pusher. A cardiac surgeon and a surgery resident each completed four manually tied rings and four COR-KNOT rings via a thoracotomy trainer. The total time to knot and cut each ring's sutures was recorded. Suture attachment pressures were measured within (intrasuture) and between (extrasuture) each suture loop using a 0.5 × 2.0-mm microtransducer probe system.

RESULTS:

The suture holding pressures for the COR-KNOT fasteners were significantly greater than for the manually tied knots (median, 1008.9 vs 415.8 mm Hg, P < 0.001). All automated fasteners measured greater than 500 mm Hg, whereas 56% of the hand-tied knots were less than 500 mm Hg, and 14% were less than 75 mm Hg. There was less variation in attachment pressures for the COR-KNOT fasteners than for the hand-tied knots (SD, 401.6 vs 499.3 mm Hg, P = 0.04). Significant time savings occurred with the use of the COR-KNOT compared with manual tying (12.4 vs 71.1 seconds per knot, P = 0.001).

CONCLUSIONS:

The novel microtransducer technology provided an innovative means of evaluating cardiac prosthetic anchoring sutures. In this model, mitral annuloplasty ring sutures secured with the COR-KNOT device were stronger, more consistent, and faster than with manually tied knots.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Automação / Suturas / Técnicas de Sutura / Procedimentos Cirúrgicos Minimamente Invasivos / Educação Médica Continuada / Anuloplastia da Valva Mitral Limite: Animals / Humans Idioma: En Revista: Innovations (Phila) Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Automação / Suturas / Técnicas de Sutura / Procedimentos Cirúrgicos Minimamente Invasivos / Educação Médica Continuada / Anuloplastia da Valva Mitral Limite: Animals / Humans Idioma: En Revista: Innovations (Phila) Ano de publicação: 2014 Tipo de documento: Article