ABSTRACT

AIM:

Objective of this study was to evaluate the anatomic changes of the stented target vessels after endovascular repair of complex aortic aneurysms.

METHODS:

Between July 2011 and December 2013, 53 aortic aneurysms were treated in our department with fenestrated and branched stent-graft devices. Forty-two of these patients were pre- and postoperatively scanned with a high resolution computer tomography (CT) (Cook Zenith® fenestrated or branched, Australia Pty. Ltd., Brisbane, Australia N.=19; AnacondaTM fenestrated, Vascutek, Glasgow, Scotland, UK N.=23). The other 11 out of the 53 patients did not receive a CT scan, because of a pre-existing renal failure. In the CT scans we retrospectively evaluated the anatomic vessel deviation at the origin of the target vessel and the vessel shift distal to the stent. For the first measurement the CT scans were loaded into OsiriX MD®, and the pre- and postoperative angles of the target vessels were measured and subtracted. For matching, the CT-scans were normalized at vertebral body lumbar 2. The second measured angle was the maximal measured angle distal to the target vessel stent-graft.

RESULTS:

Altogether, 113 target vessels were stented (celiac trunk [CT] 15, superior mesenteric arteries [SMA] 26, renal arteries [RA] 72), with 97 balloon-expandable PTFE stents 90 Atrium V12 (Maquet Getinge group, Hudson, NH, USA), 7 BeGrafts (Bentley InnoMed, Hechingen, Germany) and 16 self-expandable fluency PTFE stents (Bard, Karlsruhe, Germany). The mean anatomic deviation at the target vessel origin was 28±17.3 and the mean vessel shift distal to the stent was 36.3±18.8. There were no significant differences between the main device and the target vessel stent types.

CONCLUSION:

Fenestrated and branched stent-graft solutions for aortic aneurysm repair induce changes of the target vessel anatomy. We did not observe significant differences between the several devices.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aneurisma Aórtico / Implante de Prótese Vascular / Procedimentos Endovasculares Tipo de estudo: Observational_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: J Cardiovasc Surg (Torino) Ano de publicação: 2014 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aneurisma Aórtico / Implante de Prótese Vascular / Procedimentos Endovasculares Tipo de estudo: Observational_studies Limite: Aged / Female / Humans / Male Idioma: En Revista: J Cardiovasc Surg (Torino) Ano de publicação: 2014 Tipo de documento: Article