Radiation dose and image quality with abdominal computed tomography with automated dose-optimized tube voltage selection.
Hu, Lirong; Wang, Yujun; Hou, Hongtao; Wei, Fuquan; Yang, Guangzhao; Chen, Yougeng.
Afiliação
  • Hu L; Department of Radiology, Tongde Hospital of Zhejiang Province, Hangzhou, China.
  • Wang Y; Department of Radiology, Tongde Hospital of Zhejiang Province, Hangzhou, China 13857104269@163.com.
  • Hou H; Department of Radiology, Tongde Hospital of Zhejiang Province, Hangzhou, China.
  • Wei F; Department of Radiology, Tongde Hospital of Zhejiang Province, Hangzhou, China.
  • Yang G; Department of Radiology, Tongde Hospital of Zhejiang Province, Hangzhou, China.
  • Chen Y; Department of Radiology, Tongde Hospital of Zhejiang Province, Hangzhou, China.
J Int Med Res ; 42(4): 1011-7, 2014 Aug.
Article em En | MEDLINE | ID: mdl-24854605
ABSTRACT

OBJECTIVES:

This study assessed image quality and radiation dose of multidetector computed tomography (CT) examination using a standard protocol and a low-voltage protocol.

METHODS:

Patients requiring contrast-enhanced abdominal CT examination were randomly assigned to two groups with different voltage protocols (i) 120 kV; (ii) an automated attenuation-based tube potential optimization mode (CARE kV). The volume CT dose index (CTDIvol) and dose length product (DLP) were recorded. Image quality was semiquantitatively assessed by two blinded radiologists using a five-point scale.

RESULTS:

There were 39 patients in the 120 kV group and 50 patients in the CARE kV group. There was no obvious difference in image quality score between the groups. CARE kV resulted in a voltage reduction to 100 kV in 45 patients and to 80 kV in five patients. CTDIvol and DLP were significantly lower with CARE kV than with the 120 kV protocol.

CONCLUSIONS:

The use of CARE kV reduces radiation dose with no loss of image quality compared with a standard 120 kV protocol.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Tomografia Computadorizada por Raios X / Abdome Tipo de estudo: Clinical_trials Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Int Med Res Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Tomografia Computadorizada por Raios X / Abdome Tipo de estudo: Clinical_trials Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Int Med Res Ano de publicação: 2014 Tipo de documento: Article