Impact of study oximeter masking algorithm on titration of oxygen therapy in the Canadian oxygen trial.
Schmidt, Barbara; Roberts, Robin S; Whyte, Robin K; Asztalos, Elizabeth V; Poets, Christian; Rabi, Yacov; Solimano, Alfonso; Nelson, Harvey.
Afiliação
  • Schmidt B; Division of Neonatology, Children's Hospital of Philadelphia and University of Pennsylvania, Philadelphia, PA; Department of Clinical Epidemiology and Biostatistics, McMaster University, Hamilton, Canada. Electronic address: barbara.schmidt@uphs.upenn.edu.
  • Roberts RS; Department of Clinical Epidemiology and Biostatistics, McMaster University, Hamilton, Canada.
  • Whyte RK; Department of Pediatrics, Dalhousie University, Halifax, Canada.
  • Asztalos EV; Department of Paediatrics, University of Toronto, Toronto, Canada.
  • Poets C; Department of Neonatology, Eberhard Karls University, Tuebingen, Germany.
  • Rabi Y; Department of Pediatrics, University of Calgary, Calgary, Canada.
  • Solimano A; Department of Pediatrics, University of British Columbia, Vancouver, Canada.
  • Nelson H; Department of Clinical Epidemiology and Biostatistics, McMaster University, Hamilton, Canada.
J Pediatr ; 165(4): 666-71.e2, 2014 Oct.
Article em En | MEDLINE | ID: mdl-24973289
ABSTRACT

OBJECTIVE:

To compare oxygen saturations as displayed to caregivers on offset pulse oximeters in the 2 groups of the Canadian Oxygen Trial. STUDY

DESIGN:

In 5 double-blind randomized trials of oxygen saturation targeting, displayed saturations between 88% and 92% were offset by 3% above or below the true values but returned to true values below 84% and above 96%. During the transition, displayed values remained static at 96% in the lower and at 84% in the higher target group during a 3% change in true saturations. In contrast, displayed values changed rapidly from 88% to 84% in the lower and from 92% to 96% in the higher target group during a 1% change in true saturations. We plotted the distributions of median displayed saturations on days with >12 hours of supplemental oxygen in 1075 Canadian Oxygen Trial participants to reconstruct what caregivers observed at the bedside.

RESULTS:

The oximeter masking algorithm was associated with an increase in both stability and instability of displayed saturations that occurred during the transition between offset and true displayed values at opposite ends of the 2 target ranges. Caregivers maintained saturations at lower displayed values in the higher than in the lower target group. This differential management reduced the separation between the median true saturations in the 2 groups by approximately 3.5%.

CONCLUSIONS:

The design of the oximeter masking algorithm may have contributed to the smaller-than-expected separation between true saturations in the 2 study groups of recent saturation targeting trials in extremely preterm infants.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigênio / Oxigenoterapia / Oximetria Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Humans / Infant / Newborn País/Região como assunto: America do norte Idioma: En Revista: J Pediatr Ano de publicação: 2014 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigênio / Oxigenoterapia / Oximetria Tipo de estudo: Clinical_trials / Prognostic_studies Limite: Humans / Infant / Newborn País/Região como assunto: America do norte Idioma: En Revista: J Pediatr Ano de publicação: 2014 Tipo de documento: Article