Your browser doesn't support javascript.
loading
Veno-venous double lasso pull-and-push technique for transseptal retrieval of an embolized Watchman occluder.
Fastner, Christian; Lehmann, Ralf; Behnes, Michael; Sartorius, Benjamin; Borggrefe, Martin; Akin, Ibrahim.
Afiliação
  • Fastner C; First Department of Medicine, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany; DZHK (German Center for Cardiovascular Research) partner site Mannheim, Germany. Electronic address: christian.fastner@umm.de.
  • Lehmann R; First Department of Medicine, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany; DZHK (German Center for Cardiovascular Research) partner site Mannheim, Germany.
  • Behnes M; First Department of Medicine, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany; DZHK (German Center for Cardiovascular Research) partner site Mannheim, Germany.
  • Sartorius B; First Department of Medicine, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany; DZHK (German Center for Cardiovascular Research) partner site Mannheim, Germany.
  • Borggrefe M; First Department of Medicine, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany; DZHK (German Center for Cardiovascular Research) partner site Mannheim, Germany.
  • Akin I; First Department of Medicine, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany; DZHK (German Center for Cardiovascular Research) partner site Mannheim, Germany.
Cardiovasc Revasc Med ; 17(3): 206-8, 2016.
Article em En | MEDLINE | ID: mdl-26916569
Intraprocedural device dislodgement of a 24-mm Watchman™ left atrial appendage (LAA) closure device occurred in a 83-year-old female with a wide left atrial appendage ostium (broccoli configuration) and a surgical mitral valve reconstruction. Device rested in the atrial cavity. A second stable 12French transseptal electrophysiological sheath and two snares were needed to stabilize, elongate and gently guide the device into the second sheath. One of the snares was unclamped at its proximal end and retrogradely pulled through. After successful retrieval, a 27-mm Watchman™ device could easily be implanted in the very same session via the left delivery sheath. Patient was discharged from hospital in good general health after regular turnaround time.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibrilação Atrial / Cateterismo Cardíaco / Migração de Corpo Estranho / Apêndice Atrial / Remoção de Dispositivo / Embolia Tipo de estudo: Diagnostic_studies / Etiology_studies Limite: Aged80 / Female / Humans Idioma: En Revista: Cardiovasc Revasc Med Ano de publicação: 2016 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fibrilação Atrial / Cateterismo Cardíaco / Migração de Corpo Estranho / Apêndice Atrial / Remoção de Dispositivo / Embolia Tipo de estudo: Diagnostic_studies / Etiology_studies Limite: Aged80 / Female / Humans Idioma: En Revista: Cardiovasc Revasc Med Ano de publicação: 2016 Tipo de documento: Article