Absolute wavelength calibration of a Doppler spectrometer with a custom Fabry-Perot optical system.
Baltzer, M M; Craig, D; Den Hartog, D J; Nishizawa, T; Nornberg, M D.
Afiliação
  • Baltzer MM; Department of Physics, Wheaton College, Wheaton, Illinois 60187, USA.
  • Craig D; Department of Physics, Wheaton College, Wheaton, Illinois 60187, USA.
  • Den Hartog DJ; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA.
  • Nishizawa T; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA.
  • Nornberg MD; Department of Physics, University of Wisconsin-Madison, Madison, Wisconsin 53706, USA.
Rev Sci Instrum ; 87(11): 11E509, 2016 Nov.
Article em En | MEDLINE | ID: mdl-27910359
ABSTRACT
An Ion Doppler Spectrometer (IDS) is used for fast measurements of C VI line emission (343.4 nm) in the Madison Symmetric Torus. Absolutely calibrated flow measurements are difficult because the IDS records data within 0.25 nm of the line. Commercial calibration lamps do not produce lines in this narrow range. A light source using an ultraviolet LED and etalon was designed to provide a fiducial marker 0.08 nm wide. The light is coupled into the IDS at f/4, and a holographic diffuser increases homogeneity of the final image. Random and systematic errors in data analysis were assessed. The calibration is accurate to 0.003 nm, allowing for flow measurements accurate to 3 km/s. This calibration is superior to the previous method which used a time-averaged measurement along a chord believed to have zero net Doppler shift.
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Prognostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2016 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Prognostic_studies Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2016 Tipo de documento: Article