Spectroscopic determination of the composition of a 50 kV hydrogen diagnostic neutral beam.
Rev Sci Instrum
; 87(11): 11E543, 2016 Nov.
Article
em En
| MEDLINE
| ID: mdl-27910452
ABSTRACT
A grating spectrometer with an electron multiplying charge-coupled device camera is used to diagnose a 50 kV, 5 A, 20 ms hydrogen diagnostic neutral beam. The ion source density is determined from Stark broadened Hß emission and the spectrum of Doppler-shifted Hα emission is used to quantify the fraction of ions at full, half, and one-third beam energy under a variety of operating conditions including fueling gas pressure and arc discharge current. Beam current is optimized at low-density conditions in the ion source while the energy fractions are found to be steady over most operating conditions.