Your browser doesn't support javascript.
loading
Experimental use of a novel single-port gasless laparoendoscopic operative field formation device.
Hu, Hai; Zhang, Kai; Zhao, Gang; Jiang, Zhaoyan; Huang, Anhua; Cai, Jingli; Xu, Anan; Li, Haidong; He, Chuanqi; Ding, Kan; Lu, Ruiqi; Ye, Ruiqi.
Afiliação
  • Hu H; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Zhang K; Medical School of Tongji University, Shanghai, China.
  • Zhao G; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Jiang Z; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Huang A; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Cai J; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Xu A; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Li H; East Hospital Affiliated with Tongji University, Shanghai, China.
  • He C; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Ding K; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Lu R; East Hospital Affiliated with Tongji University, Shanghai, China.
  • Ye R; East Hospital Affiliated with Tongji University, Shanghai, China.
Ann Med Surg (Lond) ; 13: 13-19, 2017 Jan.
Article em En | MEDLINE | ID: mdl-28018588
ABSTRACT

INTRODUCTION:

Traditional laparaendoscopic surgery using CO2 pneumoperitoneum is associated with complications and the existing gasless laparaendoscopic surgery has shortcomings such as poor visibility in the operation field. To overcome the disadvantages of the current lifting gasless laparaendoscopic operation platforms, we developed an inflatable device for gasless laparoscopic operation field formation (LOFF) that can be internally installed and applied in practice.

METHODS:

We initially designed operation platforms for gasless laparaendoscopic single-port (GLESP) surgery. Subsequently, a triangular prismatic LOFF device was selected and applied successfully to GLESP cholecystectomy of five pigs. Ultimately, using pigs as a model, three surgical approaches (LOFF-assisted laparaendoscopic single-site (LOFF-LESS), LESS surgery, and traditional lifting (GLESP) were compared, and the advantages and drawbacks of inflatable devices for gasless laparoscopic operation field assessed.

RESULTS:

The use of the LOFF device in GLESP cholecystectomy was first evaluated. The time for surgical space formation (4.4 ± 1.2 and 4.8 ± 1.0), the operating time for gallbladder removal (25.2 ± 4.8 and 25.4 ± 2.7), and the loss of blood (9.4 ± 3.1and 9.2 ± 2.4) was similar between LESS and LOFF, respectively (Table 2). In contrast these parameters were higher in GLESP (6.6 ± 1.0, 30.3 ± 4.4 and 10.1 ± 2.0, respectively. The LOFF-LESS surgery exhibited a clearer exposure of the surgical field and shorter operating time than the GLESP surgery. LESS technology showed less postoperation pain, fast recovery, and extremely high cosmetic satisfaction.

CONCLUSION:

The LOFF device provides a safe, effective, and feasible operation platform that can be internally installed and inflated for GLESP surgery during cholescytectomy in animal models.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Ann Med Surg (Lond) Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Ann Med Surg (Lond) Ano de publicação: 2017 Tipo de documento: Article