Your browser doesn't support javascript.
loading
Analysis of myocardial perfusion parameters in an ex-vivo porcine heart model using third generation dual-source CT.
Pelgrim, Gert Jan; Duguay, Taylor M; Stijnen, J Marco A; Varga-Szemes, Akos; Van Tuijl, Sjoerd; Schoepf, U Joseph; Oudkerk, Matthijs; Vliegenthart, Rozemarijn.
Afiliação
  • Pelgrim GJ; University of Groningen, University Medical Center Groningen, Center for Medical Imaging - North East Netherlands, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.
  • Duguay TM; Medical University of South Carolina, Department of Radiology, 25 Courtenay Drive, 29425 SC, Charleston, SC, USA.
  • Stijnen JM; LifeTec Group BV, Kennedyplein 10-11, 5611 ZS, Eindhoven, The Netherlands.
  • Varga-Szemes A; Medical University of South Carolina, Department of Radiology, 25 Courtenay Drive, 29425 SC, Charleston, SC, USA.
  • Van Tuijl S; LifeTec Group BV, Kennedyplein 10-11, 5611 ZS, Eindhoven, The Netherlands.
  • Schoepf UJ; Medical University of South Carolina, Department of Radiology, 25 Courtenay Drive, 29425 SC, Charleston, SC, USA.
  • Oudkerk M; University of Groningen, University Medical Center Groningen, Center for Medical Imaging - North East Netherlands, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands.
  • Vliegenthart R; University of Groningen, University Medical Center Groningen, Center for Medical Imaging - North East Netherlands, Hanzeplein 1, 9713 GZ, Groningen, The Netherlands; University of Groningen, University Medical Center Groningen, Department of Radiology, Hanzeplein 1, 9713 GZ, Groningen, The Netherlan
J Cardiovasc Comput Tomogr ; 11(2): 141-147, 2017.
Article em En | MEDLINE | ID: mdl-28202246
ABSTRACT

PURPOSE:

To evaluate the relationship between fractional flow reserve (FFR)-determined coronary artery stenosis severity and myocardial perfusion parameters derived from dynamic myocardial CT perfusion imaging (CTP) in an ex-vivo porcine heart model.

METHODS:

Six porcine hearts were perfused according to Langendorff. Circulatory parameters such as arterial blood flow (ABF) (L/min), mean arterial pressure (MAP) (mmHg) and heart rate (bpm) were monitored. Using an inflatable cuff and monitored via a pressure wire, coronary artery stenoses of different FFR grades were created (no stenosis, FFR = 0.80, FFR = 0.70, FFR = 0.60, and FFR = 0.50). Third generation dual-source CT was used to perform dynamic CTP in shuttle mode at 70 kV. Using the AHA-16-segment model, myocardial blood flow (MBF) (mL/100 mL/min) and volume (MBV) (mL/100 mL) were analyzed using dedicated software for all ischaemic and non-ischaemic segments.

RESULTS:

During five successful experiments, ABF ranged from 0.8 to 1.2 L/min, MAP from 73 to 90 mmHg and heart rate from 83 to 115 bpm. Non-ischaemic and ischaemic segments showed significant differences in MBF for stenosis grades of FFR ≤ 0.70. At this degree of obstruction, median MBF was 79 (interquartile range [IQR] 66-90) for non-ischaemic segments versus 56 mL/100 mL/min (IQR 46-73) for ischaemic segments (p < 0.05). For MBV, a significant difference was found at FFR ≤ 0.80 with median MBV values of 7.6 (IQR 7.0-8.3) and 7.1 mL/100 mL (IQR 6.0-8.2) for non-ischaemic and ischaemic myocardial segments, respectively (p < 0.05).

CONCLUSION:

Artificial flow alterations in a Langendorff porcine heart model could be detected and measured by CTP-derived myocardial perfusion parameters and showed significant systematic correlation with stepwise flow reduction that permitted early detection of ischaemic myocardium. Additional research in clinical setting is required to develop absolute quantitative CTP.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angiografia Coronária / Vasos Coronários / Estenose Coronária / Reserva Fracionada de Fluxo Miocárdico / Imagem de Perfusão do Miocárdio / Angiografia por Tomografia Computadorizada Tipo de estudo: Prognostic_studies / Screening_studies Limite: Animals Idioma: En Revista: J Cardiovasc Comput Tomogr Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Angiografia Coronária / Vasos Coronários / Estenose Coronária / Reserva Fracionada de Fluxo Miocárdico / Imagem de Perfusão do Miocárdio / Angiografia por Tomografia Computadorizada Tipo de estudo: Prognostic_studies / Screening_studies Limite: Animals Idioma: En Revista: J Cardiovasc Comput Tomogr Ano de publicação: 2017 Tipo de documento: Article