Oesophageal heat exchangers with a diameter of 11mm or 14.7mm are equally effective and safe for targeted temperature management.
Schroeder, Daniel C; Guschlbauer, Maria; Maul, Alexandra C; Cremer, Daniel A; Becker, Ingrid; de la Puente Bethencourt, David; Paal, Peter; Padosch, Stephan A; Wetsch, Wolfgang A; Annecke, Thorsten; Böttiger, Bernd W; Sterner-Kock, Anja; Herff, Holger.
Afiliação
  • Schroeder DC; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Guschlbauer M; Department of Experimental Medicine, University Hospital of Cologne, Cologne, Germany.
  • Maul AC; Department of Experimental Medicine, University Hospital of Cologne, Cologne, Germany.
  • Cremer DA; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Becker I; Institute of Medical Statistics, Informatics and Epidemiology, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • de la Puente Bethencourt D; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Paal P; Barts Heart Centre, St Bartholomew's Hospital, Barts Health NHS Trust. Queen Mary University of London, London, United Kingdom.
  • Padosch SA; Barmherzige Brüder Salzburg Hospital, Department of Anaesthesiology and Critical Care Medicine, Kajetanerplatz 2, Salzburg, Austria.
  • Wetsch WA; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Annecke T; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Böttiger BW; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Sterner-Kock A; Department of Anaesthesiology and Intensive Care Medicine, University Hospital of Cologne, Kerpener Straße 62, Cologne, Germany.
  • Herff H; Department of Experimental Medicine, University Hospital of Cologne, Cologne, Germany.
PLoS One ; 12(3): e0173229, 2017.
Article em En | MEDLINE | ID: mdl-28291783
ABSTRACT

BACKGROUND:

Targeted temperature management (TTM) is widely used in critical care settings for conditions including hepatic encephalopathy, hypoxic ischemic encephalopathy, meningitis, myocardial infarction, paediatric cardiac arrest, spinal cord injury, traumatic brain injury, ischemic stroke and sepsis. Furthermore, TTM is a key treatment for patients after out-of-hospital cardiac-arrest (OHCA). However, the optimal cooling method, which is quick, safe and cost-effective still remains controversial. Since the oesophagus is adjacent to heart and aorta, fast heat-convection to the central blood-stream could be achieved with a minimally invasive oesophageal heat exchanger (OHE). To date, the optimal diameter of an OHE is still unknown. While larger diameters may cause thermal- or pressure-related tissue damage after long-term exposure to the oesophageal wall, smaller diameter (e.g., gastric tubes, up to 11mm) may not provide effective cooling rates. Thus, the objective of the study was to compare OHE-diameters of 11mm (OHE11) and 14.7mm (OHE14.7) and their effects on tissue and cooling capability.

METHODS:

Pigs were randomized to OHE11 (N = 8) or OHE14.7 (N = 8). After cooling, pigs were maintained at 33°C for 1 hour. After 10h rewarming, oesophagi were analyzed by means of histopathology. The oesophagus of four animals from a separate study that underwent exactly the identical preparation and cooling protocol described above but received a maintenance period of 24h were used as histopathological controls.

RESULTS:

Mean cooling rates were 2.8±0.4°C°C/h (OHE11) and 3.0±0.3°C °C/h (OHE14.7; p = 0.20). Occasional mild acute inflammatory transepithelial infiltrates were found in the cranial segment of the oesophagus in all groups including controls. Deviations from target temperature were 0.1±0.4°C (OHE11) and 0±0.1°C (OHE14.7; p = 0.91). Rewarming rates were 0.19±0.07°C °C/h (OHE11) and 0.20±0.05°C °C/h (OHE14.7; p = 0.75).

CONCLUSIONS:

OHE with diameters of 11 mm and 14.7 mm achieve effective cooling rates for TTM and did not cause any relevant oesophageal tissue damage. Both OHE demonstrated acceptable deviations from target temperature and allowed for an intended rewarming rate (0.25°C/h).
Assuntos

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 4_TD Base de dados: MEDLINE Assunto principal: Temperatura Corporal / Temperatura Baixa / Esôfago / Temperatura Alta Tipo de estudo: Guideline Limite: Animals Idioma: En Revista: PLoS One Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 4_TD Base de dados: MEDLINE Assunto principal: Temperatura Corporal / Temperatura Baixa / Esôfago / Temperatura Alta Tipo de estudo: Guideline Limite: Animals Idioma: En Revista: PLoS One Ano de publicação: 2017 Tipo de documento: Article