Calculation of the time resolution of the J-PET tomograph using kernel density estimation.
Phys Med Biol ; 62(12): 5076-5097, 2017 Jun 21.
Article em En | MEDLINE | ID: mdl-28452337
ABSTRACT
In this paper we estimate the time resolution of the J-PET scanner built from plastic scintillators. We incorporate the method of signal processing using the Tikhonov regularization framework and the kernel density estimation method. We obtain simple, closed-form analytical formulae for time resolution. The proposed method is validated using signals registered by means of the single detection unit of the J-PET tomograph built from a 30 cm long plastic scintillator strip. It is shown that the experimental and theoretical results obtained for the J-PET scanner equipped with vacuum tube photomultipliers are consistent.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Contagem de Cintilação / Tomografia por Emissão de Pósitrons Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 2017 Tipo de documento: Article