Your browser doesn't support javascript.
loading
Autoregressive Modeling of Drift and Random Error to Characterize a Continuous Intravascular Glucose Monitoring Sensor.
Zhou, Tony; Dickson, Jennifer L; Geoffrey Chase, J.
Afiliação
  • Zhou T; 1 Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand.
  • Dickson JL; 1 Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand.
  • Geoffrey Chase J; 1 Department of Mechanical Engineering, University of Canterbury, Christchurch, New Zealand.
J Diabetes Sci Technol ; 12(1): 90-104, 2018 01.
Article em En | MEDLINE | ID: mdl-28707484
ABSTRACT

BACKGROUND:

Continuous glucose monitoring (CGM) devices have been effective in managing diabetes and offer potential benefits for use in the intensive care unit (ICU). Use of CGM devices in the ICU has been limited, primarily due to the higher point accuracy errors over currently used traditional intermittent blood glucose (BG) measures. General models of CGM errors, including drift and random errors, are lacking, but would enable better design of protocols to utilize these devices. This article presents an autoregressive (AR) based modeling method that separately characterizes the drift and random noise of the GlySure CGM sensor (GlySure Limited, Oxfordshire, UK).

METHODS:

Clinical sensor data (n = 33) and reference measurements were used to generate 2 AR models to describe sensor drift and noise. These models were used to generate 100 Monte Carlo simulations based on reference blood glucose measurements. These were then compared to the original CGM clinical data using mean absolute relative difference (MARD) and a Trend Compass.

RESULTS:

The point accuracy MARD was very similar between simulated and clinical data (9.6% vs 9.9%). A Trend Compass was used to assess trend accuracy, and found simulated and clinical sensor profiles were similar (simulated trend index 11.4° vs clinical trend index 10.9°).

CONCLUSION:

The model and method accurately represents cohort sensor behavior over patients, providing a general modeling approach to any such sensor by separately characterizing each type of error that can arise in the data. Overall, it enables better protocol design based on accurate expected CGM sensor behavior, as well as enabling the analysis of what level of each type of sensor error would be necessary to obtain desired glycemic control safety and performance with a given protocol.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 Base de dados: MEDLINE Assunto principal: Glicemia / Automonitorização da Glicemia / Diabetes Mellitus Tipo de estudo: Clinical_trials / Guideline / Health_economic_evaluation Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Diabetes Sci Technol Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 Base de dados: MEDLINE Assunto principal: Glicemia / Automonitorização da Glicemia / Diabetes Mellitus Tipo de estudo: Clinical_trials / Guideline / Health_economic_evaluation Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Diabetes Sci Technol Ano de publicação: 2018 Tipo de documento: Article