A compact membrane-driven diamond anvil cell and cryostat system for nuclear resonant scattering at high pressure and low temperature.
Zhao, J Y; Bi, W; Sinogeikin, S; Hu, M Y; Alp, E E; Wang, X C; Jin, C Q; Lin, J F.
Afiliação
  • Zhao JY; Advanced Photon Source, Argonne National Laboratory, 9700 S Cass Ave., Argonne, Illinois 60439, USA.
  • Bi W; Advanced Photon Source, Argonne National Laboratory, 9700 S Cass Ave., Argonne, Illinois 60439, USA.
  • Sinogeikin S; High Pressure Collaborative Access Team, Geophysical Laboratory, Carnegie Institution of Washington, Argonne, Illinois 60439, USA.
  • Hu MY; Advanced Photon Source, Argonne National Laboratory, 9700 S Cass Ave., Argonne, Illinois 60439, USA.
  • Alp EE; Advanced Photon Source, Argonne National Laboratory, 9700 S Cass Ave., Argonne, Illinois 60439, USA.
  • Wang XC; Institute of Physics, Chinese Academy of Sciences, Beijing 10090, People's Republic of China.
  • Jin CQ; Institute of Physics, Chinese Academy of Sciences, Beijing 10090, People's Republic of China.
  • Lin JF; Department of Geology Sciences, The University of Texas at Austin, Austin, Texas 78712, USA.
Rev Sci Instrum ; 88(12): 125109, 2017 Dec.
Article em En | MEDLINE | ID: mdl-29289218
ABSTRACT
A new miniature panoramic diamond anvil cell (mini-pDAC) as well as a unique gas membrane-driven mechanism is developed and implemented to measure electronic, magnetic, vibrational, and thermodynamic properties of materials using the nuclear resonant inelastic X-ray scattering (NRIXS) and the synchrotron Mössbauer spectroscopy (SMS) simultaneously at high pressure (over Mbar) and low temperature (T < 10 K). The gas membrane system allows in situ pressure tuning of the mini-pDAC at low temperature. The mini-pDAC fits into a specially designed compact liquid helium flow cryostat system to achieve low temperatures, where liquid helium flows through the holder of the mini-pDAC to cool the sample more efficiently. The system has achieved sample temperatures as low as 9 K. Using the membrane, sample pressures of up to 1.4 Mbar have been generated from this mini-pDAC. The instrument has been routinely used at 3-ID, Advanced Photon Source, for NRIXS and SMS studies. The same instrument can easily be used for other X-ray techniques, such as X-ray radial diffraction, X-ray Raman scattering, X-ray emission spectroscopy, and X-ray inelastic scattering under high pressure and low temperature. In this paper, technical details of the mini-pDAC, membrane engaging mechanism, and the cryostat system are described, and some experimental results are discussed.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2017 Tipo de documento: Article