A high-throughput headspace gas chromatographic technique for the determination of nitrite content in water samples.
J Chromatogr A
; 1538: 104-107, 2018 Feb 23.
Article
em En
| MEDLINE
| ID: mdl-29361281
This paper reports on a high-throughput headspace gas chromatographic method (HS-GC) for the determination of nitrite content in water sample, based on GC measurement of cyclohexene produced from the reaction between nitrite and cyclamate in a closed vial. The method has a relative standard deviation of <3.5%; The differences between the results of the nitrite measurements obtained by this method and those of a reference method were less than 5.8% and the recoveries of the method were in the range of 94.8-102% (for a spiked nitrite content range from 0.002 to 0.03â¯mg/L). The limit of detection of the method was 0.46⯵gâ¯L-1. Due to an overlapping mode in the headspace auto-sampler system, the method can provide an automated and high-throughput nitrite analysis for the surface water samples. In short, the present HS-GC method is simple, accurate, and sensitive, and it is very suitable to be used in the batch sample testing.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Contexto em Saúde:
12_ODS3_hazardous_contamination
/
2_ODS3
Base de dados:
MEDLINE
Assunto principal:
Monitoramento Ambiental
/
Cromatografia Gasosa
/
Água Doce
/
Nitritos
Idioma:
En
Revista:
J Chromatogr A
Ano de publicação:
2018
Tipo de documento:
Article