Label-free and simple detection of endotoxins using a sensitive LSPR biosensor based on silver nanocolumns.
Anal Biochem
; 548: 96-101, 2018 05 01.
Article
em En
| MEDLINE
| ID: mdl-29501578
This paper describes the construction of a silver-based LSPR biosensor for endotoxin detection. We used GLAD method to procure reproducible silver nanocolumns. In this work, the silver nanostructures were considerably stabilized by a SAM of MPA, and the limit of detection of biosensor was measured to be 340â¯pg/ml for endotoxin E. coli. Considering endotoxin B. abortus as the second type of endotoxin contamination in our target samples (HBs-ag produced in Institute Pasteur, Iran), we investigated selectivity of the biosensor in various experiments. We showed that this biosensor can selectively detect both types of endotoxins compared to other biological species. Overall, this study proposes that LSPR biosensing can be considered as a sensitive, simple, and label-free method for endotoxin detection in the quality control laboratories.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Contexto em Saúde:
3_ND
Base de dados:
MEDLINE
Assunto principal:
Prata
/
Técnicas Biossensoriais
/
Endotoxinas
/
Escherichia coli
/
Nanopartículas Metálicas
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Revista:
Anal Biochem
Ano de publicação:
2018
Tipo de documento:
Article