Your browser doesn't support javascript.
loading
Simulation of the Effect of Different Presbyopia-Correcting Intraocular Lenses With Eyes With Previous Laser Refractive Surgery.
J Refract Surg ; 34(4): 222-227, 2018 Apr 01.
Article em En | MEDLINE | ID: mdl-29634836
ABSTRACT

PURPOSE:

To simulate the optical performance of three presbyopia-correcting intraocular lenses (IOLs) implanted in eyes with previous laser refractive surgery.

METHODS:

A simulation of the through-focus modulation transfer function (MTF) was performed for three presbyopia-correcting IOLs (Mplus, Oculentis GmbH, Berlin, Germany; Symfony, Johnson & Johnson Vision, Santa Ana, CA; and Mini Well, SIFI S.p.A., Lavinaio, Italy) in one eye with previous myopic LASIK and another with hyperopic LASIK. Real topographic data and the wavefront aberration profile of each IOL obtained with a Hartmann-Shack sensor were used.

RESULTS:

In the eye with myopic LASIK, all IOLs lost optical quality at near and intermediate distances for 4- and 4.7-mm pupil size. For 3-mm pupil size, the Mini Well IOL showed the best intermediate and near MTF and maintained the far focus independently of the pupil. In the eye with hyperopic LASIK, the Mini Well IOL showed an intermediate, distance, and -4.00-diopter (D) foci for all pupils. The Symfony IOL showed a depth of focus at far and intermediate distance for 3-mm and a focus at -2.50 D in the rest. The Mplus showed a focus of -4.50 and -3.00 D for the 3- and 4-mm pupil, respectively.

CONCLUSIONS:

The Mini Well and Symfony IOLs seem to work better than the Mplus IOL in eyes with previous myopic LASIK. With previous hyperopic LASIK, the Mini Well IOL seems to be able to provide acceptable near, intermediate, and far foci for all pupil sizes. These findings should be confirmed in future clinical studies. [J Refract Surg. 2018;34(4)222-227.].
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Presbiopia / Ceratomileuse Assistida por Excimer Laser In Situ / Lasers de Excimer / Óptica e Fotônica / Hiperopia / Lentes Intraoculares / Miopia Tipo de estudo: Prognostic_studies Limite: Adult / Humans / Male Idioma: En Revista: J Refract Surg Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Presbiopia / Ceratomileuse Assistida por Excimer Laser In Situ / Lasers de Excimer / Óptica e Fotônica / Hiperopia / Lentes Intraoculares / Miopia Tipo de estudo: Prognostic_studies Limite: Adult / Humans / Male Idioma: En Revista: J Refract Surg Ano de publicação: 2018 Tipo de documento: Article