Hearing outcomes of the active bone conduction system Bonebridge® in conductive or mixed hearing loss. / Resultados auditivos del sistema activo de conducción ósea Bonebridge® en hipoacusias conductivas o mixtas.
Carnevale, Claudio; Til-Pérez, Guillermo; Arancibia-Tagle, Diego J; Tomás-Barberán, Manuel D; Sarría-Echegaray, Pedro L.
Afiliação
  • Carnevale C; Servicio de Otorrinolaringología y Cirugía de Cabeza y Cuello, Hospital Universitario Son Espases, Palma de Mallorca, España. Electronic address: claudio.carnevale.orl@gmail.com.
  • Til-Pérez G; Servicio de Otorrinolaringología y Cirugía de Cabeza y Cuello, Hospital Universitario Son Espases, Palma de Mallorca, España.
  • Arancibia-Tagle DJ; Servicio de Otorrinolaringología y Cirugía de Cabeza y Cuello, Hospital Universitario Son Espases, Palma de Mallorca, España.
  • Tomás-Barberán MD; Servicio de Otorrinolaringología y Cirugía de Cabeza y Cuello, Hospital Universitario Son Espases, Palma de Mallorca, España.
  • Sarría-Echegaray PL; Servicio de Otorrinolaringología y Cirugía de Cabeza y Cuello, Hospital Universitario Son Espases, Palma de Mallorca, España.
Article em En, Es | MEDLINE | ID: mdl-29784242
ABSTRACT

OBJECTIVE:

The active transcutaneous bone conduction implant Bonebridge®, is indicated for patients affected by bilateral conductive/mixed hearing loss or unilateral sensorineural hearing loss, showing hearing outcomes similar to other percutaneous bone conduction implants, but with a lower rate of complications. The aim of this study was to analyze the hearing outcomes in a series of 26 patients affected by conductive or mixed hearing loss and treated with Bonebridge®.

METHODS:

26 of 30 patients implanted with Bonebridge® between October 2012 and May 2017, were included in the study. We compared the air conduction thresholds at the frequencies 500, 1000, 2000, 3000, 4000Hz, the SRT50% and the percentage of correct answers at an intensity of 50dB with and without the implant.

RESULTS:

"Pure tone average" with the implant was 34.91dB showing an average gain of 33.46dB. Average SRT 50% with the implant was 34.33dB, whereas before the surgery no patient achieved 50% of correct answers at a sound intensity of 50dB. The percentage of correct answers at 50dB changed from 11% without the implant to 85% with it. We only observed one complication consisting of an extrusion of the implant in a patient with a history of 2 previous rhytidectomies.

CONCLUSIONS:

The hearing outcomes obtained in our study are similar to those published in the literature. Bonebridge® represents an excellent alternative in the treatment of conductive or mixed hearing loss, and with a lower rate of complications.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Condução Óssea / Perda Auditiva Condutiva-Neurossensorial Mista / Prótese Ancorada no Osso / Perda Auditiva Bilateral Tipo de estudo: Evaluation_studies / Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En / Es Revista: Acta Otorrinolaringol Esp (Engl Ed) Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Condução Óssea / Perda Auditiva Condutiva-Neurossensorial Mista / Prótese Ancorada no Osso / Perda Auditiva Bilateral Tipo de estudo: Evaluation_studies / Observational_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En / Es Revista: Acta Otorrinolaringol Esp (Engl Ed) Ano de publicação: 2019 Tipo de documento: Article