Your browser doesn't support javascript.
loading
Distinguishing Highly Asymmetric Keratoconus Eyes Using Combined Scheimpflug and Spectral-Domain OCT Analysis.
Hwang, Eric S; Perez-Straziota, Claudia E; Kim, Sang Woo; Santhiago, Marcony R; Randleman, J Bradley.
Afiliação
  • Hwang ES; Keck School of Medicine of the University of Southern California, Los Angeles, California.
  • Perez-Straziota CE; USC Roski Eye Institute, Los Angeles, California.
  • Kim SW; Department of Ophthalmology, Ulsan University of College of Medicine, Ulsan, South Korea.
  • Santhiago MR; USC Roski Eye Institute, Los Angeles, California; Department of Ophthalmology at University of Sao Paulo, Sao Paulo, Brazil; Department of Ophthalmology at Federal University of Rio de Janeiro, Sao Paulo, Brazil.
  • Randleman JB; Keck School of Medicine of the University of Southern California, Los Angeles, California; USC Roski Eye Institute, Los Angeles, California. Electronic address: randlema@usc.edu.
Ophthalmology ; 125(12): 1862-1871, 2018 12.
Article em En | MEDLINE | ID: mdl-30055838
ABSTRACT

PURPOSE:

To determine optimal objective, machine-derived variables and variable combinations from Scheimpflug and spectral-domain (SD) OCT imaging to distinguish the clinically unaffected eye in patients with asymmetric keratoconus (AKC) from a normal control population.

DESIGN:

Retrospective case-control study.

PARTICIPANTS:

Thirty clinically unaffected eyes with no physical findings on slit-lamp examination, no definitive abnormalities on corneal imaging, and corrected distance acuity of 20/20 or better from 30 patients with highly AKC eyes and 60 eyes from 60 normal control patients who had undergone uneventful LASIK with at least 2 years of stable follow-up (controls).

METHODS:

Scheimpflug and SD OCT imaging were obtained in all eyes, and receiver operating characteristic (ROC) curves were generated to determine area under the curve (AUC), sensitivity, and specificity for each machine-derived variable and variable combination. MAIN OUTCOME

MEASURES:

Distinguishing AKC eyes from controls as determined by AUC, sensitivity, and specificity.

RESULTS:

No individual machine-derived metric from Scheimpflug or SD OCT technology yielded an AUC higher than 0.75. Combining 5 Scheimpflug metrics (index height decentration [IHD], index vertical asymmetry [IVA], pachymetry apex, inferior-superior value, and Ambrosio's Relational Thickness Maximum [ARTmax]) yielded the best Scheimpflug results (AUC 0.86, sensitivity 83%, specificity 83%). Combining 11 SD OCT thickness metrics (minimum-median, temporal outer, superior nasal outer, minimum, epithelium minimum-maximum, epithelial standard deviation, superior inner, superior outer, superior temporal outer, superior nasal inner, central) yielded the best SD OCT results (AUC 0.96, sensitivity 89%, specificity 89%). Combining 13 total Scheimpflug/SD OCT metrics yielded the best results overall (AUC 1.0, sensitivity 100%, specificity 100%). The most impactful variables in combined models included epithelial thickness variability and total focal corneal thickness variability from SD OCT and anterior curvature and topometric indices from Scheimpflug technology. No posterior corneal metrics were impactful in modeling.

CONCLUSIONS:

Individual machine-derived metrics from Scheimpflug and SD OCT imaging poorly distinguished normal eyes from minimally affected eyes from patients with highly AKC. Combined SD OCT metrics performed better than combined Scheimpflug metrics. Combining anterior curvature and asymmetry indices from Scheimpflug with regional total thickness and epithelial thickness variability metrics from SD OCT clearly distinguished the 2 populations. Posterior corneal indices were not useful in distinguishing populations.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Córnea / Topografia da Córnea / Tomografia de Coerência Óptica / Ceratocone Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Ophthalmology Ano de publicação: 2018 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Córnea / Topografia da Córnea / Tomografia de Coerência Óptica / Ceratocone Tipo de estudo: Observational_studies / Prognostic_studies Limite: Adolescent / Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Ophthalmology Ano de publicação: 2018 Tipo de documento: Article