Your browser doesn't support javascript.
loading
Clinical impact of cardiovascular magnetic resonance with optimized myocardial scar detection in patients with cardiac implantable devices.
Bhuva, Anish N; Kellman, Peter; Graham, Adam; Ramlall, Manish; Boubertakh, Redha; Feuchter, Patricia; Hawkins, Angela; Lowe, Martin; Lambiase, Pier D; Sekhri, Neha; Schilling, Richard J; Moon, James C; Manisty, Charlotte H.
Afiliação
  • Bhuva AN; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK; Institutes for Cardiovascular Science, University College London, London, UK.
  • Kellman P; National Institutes of Health, Bethesda, USA.
  • Graham A; Department of Cardiac Electrophysiology, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Ramlall M; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Boubertakh R; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Feuchter P; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Hawkins A; Department of Cardiac Electrophysiology, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Lowe M; Department of Cardiac Electrophysiology, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Lambiase PD; Institutes for Cardiovascular Science, University College London, London, UK; Department of Cardiac Electrophysiology, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Sekhri N; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Schilling RJ; Department of Cardiac Electrophysiology, Barts Heart Centre, St Bartholomew's Hospital, London, UK.
  • Moon JC; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK; Institutes for Cardiovascular Science, University College London, London, UK.
  • Manisty CH; Department of Cardiovascular Imaging, Barts Heart Centre, St Bartholomew's Hospital, London, UK; Institutes for Cardiovascular Science, University College London, London, UK. Electronic address: c.manisty@ucl.ac.uk.
Int J Cardiol ; 279: 72-78, 2019 Mar 15.
Article em En | MEDLINE | ID: mdl-30642645
ABSTRACT

BACKGROUND:

Myocardial scar assessment using late gadolinium enhancement Cardiovascular Magnetic Resonance (LGE CMR) is commonly indicated for patients with cardiac implantable electronic devices (CIEDs), however metal artifact can degrade images. We evaluated the clinical impact of LGE CMR incorporating a device-dependent metal artifact reduction strategy in patients with CIEDs.

METHODS:

136 CMR studies were performed in 133 consecutive patients (age 56 ±â€¯19 years, 69% male) with CIEDs (22% implantable loop recorders [ILRs], 40% permanent pacemakers [PPMs], 38% implantable cardioverter defibrillators [ICDs]; 42% non-MRI conditional) over 2 years, without complication. LGE imaging was tailored to the CIED, using a wideband sequence for left-sided PPMs and ICDs and conventional sequences for ILRs and right-sided PPMs, scoring segmental artifact. Diagnostic utility and impact on clinical management were scored by consensus of experts.

RESULTS:

CMR provided unexpected diagnoses in 22 (16%) and changed management in 113 (83%) patients. Myocardial scar was present in 92 (68%), with other abnormalities detected in another 13%. Using conventional LGE, 43 (32%) studies were non-diagnostic (79% of defibrillators) compared to 0% using wideband LGE imaging. Wideband LGE results changed clinical management in an additional 39 (75%) defibrillator patients and 10 (19%) pacemaker patients when compared to imaging with conventional LGE sequences.

CONCLUSION:

The clinical yield from CMR using optimized LGE sequences in patients with CIEDs is high with no demonstrated clinical risk. A device-dependent LGE imaging strategy using wideband LGE is needed to achieve clinical utility especially in ICD recipients.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Cicatriz / Desfibriladores Implantáveis / Imagem Cinética por Ressonância Magnética / Miocárdio Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Cardiol Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Cicatriz / Desfibriladores Implantáveis / Imagem Cinética por Ressonância Magnética / Miocárdio Tipo de estudo: Clinical_trials / Diagnostic_studies / Etiology_studies Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Cardiol Ano de publicação: 2019 Tipo de documento: Article