Your browser doesn't support javascript.
loading
Infections and outcomes after cardiac surgery-The impact of outbreaks traced to transesophageal echocardiography probes.
Vesteinsdottir, Edda; Helgason, Kristjan Orri; Sverrisson, Kristinn Orn; Gudlaugsson, Olafur; Karason, Sigurbergur.
Afiliação
  • Vesteinsdottir E; Department of Anaesthesia and Intensive Care, Landspitali - The National University Hospital of Iceland, Reykjavik, Iceland.
  • Helgason KO; Department of Clinical Microbiology, Landspitali - The National University Hospital of Iceland, Reykjavik, Iceland.
  • Sverrisson KO; Department of Anaesthesia and Intensive Care, Landspitali - The National University Hospital of Iceland, Reykjavik, Iceland.
  • Gudlaugsson O; Department of Infectious Diseases, Landspitali - The National University Hospital of Iceland, Reykjavik, Iceland.
  • Karason S; Department of Anaesthesia and Intensive Care, Landspitali - The National University Hospital of Iceland, Reykjavik, Iceland.
Acta Anaesthesiol Scand ; 63(7): 871-878, 2019 08.
Article em En | MEDLINE | ID: mdl-30888057
ABSTRACT

BACKGROUND:

Infections are a frequent complication of cardiac surgery. The intraoperative use of transesophageal echocardiography (TEE) may be an underrecognized risk factor for post-operative infections. The aim of this study was to investigate infection rates and outcomes after cardiac surgery in a nationwide cohort, especially in relation to periods where surface damaged TEE probes were used.

METHODS:

This was a retrospective, observational study at Landspitali University Hospital. All consecutive cardiac surgery patients from 1 January 2013 to 31 December 2017 were included. Patients' charts were reviewed for evidence of infection, post-operative complications or death.

RESULTS:

During the study period, 973 patients underwent cardiac surgery at Landspitali and 198 (20.3%) developed a post-operative infection. The most common infections were Pneumonia (9.1%), superficial surgical site (5.7%), bloodstream (2.8%) and deep sternal wound (1.7%). Risk factors for developing an infection included The duration of procedure, age, insulin-dependent diabetes, EuroScore II, reoperation for bleeding and an operation in a period with a surface damaged TEE probe in use. Twenty-two patients were infected with a multidrug resistant strain of Klebsiella oxytoca, 10 patients with Pseudomonas aeruginosa and two patients developed endocarditis with Enterococcus faecalis. All three pathogens were cultured from the TEE probe in use at respective time, after decontamination. The 30-day mortality rate in the patient cohort was 3.2%.

CONCLUSIONS:

The intraoperative use of surface damaged TEE probes caused two serious infection outbreaks in patients after cardiac surgery. TEE probes need careful visual inspection during decontamination and probe sheaths are recommended.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Infecção da Ferida Cirúrgica / Ecocardiografia Transesofagiana / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Acta Anaesthesiol Scand Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Complicações Pós-Operatórias / Infecção da Ferida Cirúrgica / Ecocardiografia Transesofagiana / Procedimentos Cirúrgicos Cardíacos Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Acta Anaesthesiol Scand Ano de publicação: 2019 Tipo de documento: Article