Your browser doesn't support javascript.
loading
Comparison between optical microscopy and automation for cytometric analysis of pericardial fluids in a cohort of adult subjects undergoing cardiac surgery.
Buoro, Sabrina; Seghezzi, Michela; Baigorria Vaca, Maria Del Carmen; Manenti, Barbara; Moioli, Valentina; Previtali, Giulia; Simon, Caterina; Cugola, Diego; Brucato, Antonio; Ottomano, Cosimo; Lippi, Giuseppe.
Afiliação
  • Buoro S; Clinical Chemistry Laboratory, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Seghezzi M; Clinical Chemistry Laboratory, ASST Papa Giovanni XXIII, Bergamo, Italy m.seghezzi89@gmail.com.
  • Baigorria Vaca MDC; Clinical Chemistry Laboratory, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Manenti B; Clinical Chemistry Laboratory, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Moioli V; Clinical Chemistry Laboratory, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Previtali G; Clinical Chemistry Laboratory, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Simon C; Cardiac Surgery, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Cugola D; Cardiac Surgery, ASST Papa Giovanni XXIII, Bergamo, Italy.
  • Brucato A; Internal Medicine Division, ASST Fatebenefratelli-Sacco, Milano, Italy.
  • Ottomano C; Clinical Chemistry Laboratory, SYNLAB Castenedolo, Castenedolo, Italy.
  • Lippi G; Section of Clinical Biochemistry, University of Verona, Verona, Italy.
J Clin Pathol ; 72(7): 493-500, 2019 Jul.
Article em En | MEDLINE | ID: mdl-31036703
ABSTRACT

AIMS:

Limited information is available on number and type of cells present in the pericardial fluid (PF). Current evidence and has been garnered with inaccurate application of guidelines for analysis of body fluids. This study was aimed at investigating the performance of automate cytometric analysis of PF in adult subjects.

METHODS:

Seventy-four consecutive PF samples were analysed with Sysmex XN with a module for body fluid analysis (XN-BF) and optical microscopy (OM). The study also encompassed the assessment of limit of blank, limit of detection and limit of quantitation (LoQ), imprecision, carryover and linearity of XN-BF module.

RESULTS:

XN-BF parameters were compared with OM for the following cell classes total cells (TC), leucocytes (white blood cell [WBC]), polymorphonuclear (PMN) and mononuclear (MN) cells. The relative bias were -4.5%, 71.2%, 108.2% and -47.7%, respectively. Passing and Bablok regression yielded slope comprised between 0.06 for MN and 5.8 for PMN, and intercept between 0.7 for PMN and 220.3 for MN. LoQ was comprised between 3.8×106 and 6.0×106 cells/L for WBC and PMN. Linearity was acceptable and carryover negligible.

CONCLUSIONS:

PF has a specific cellular composition. Overall, automated cell counting can only be suggested for total number of cells, whereas OM seems still the most reliable option for cell differentiation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Líquido Pericárdico Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male Idioma: En Revista: J Clin Pathol Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Líquido Pericárdico Tipo de estudo: Etiology_studies / Incidence_studies / Observational_studies / Risk_factors_studies Limite: Female / Humans / Male Idioma: En Revista: J Clin Pathol Ano de publicação: 2019 Tipo de documento: Article