Your browser doesn't support javascript.
loading
Compact Grating-Coupled Biosensor for the Analysis of Thrombin.
Kotlarek, Daria; Vorobii, Mariia; Ogieglo, Wojciech; Knoll, Wolfgang; Rodriguez-Emmenegger, Cesar; Dostálek, Jakub.
Afiliação
  • Kotlarek D; Biosensor Technologies , AIT-Austrian Institute of Technology GmbH , Konrad-Lorenz-Straße 24 , 3430 Tulln an der Donau , Austria.
  • Vorobii M; DWI - Leibniz Institute for Interactive Materials and Institute of Technical and Macromolecular Chemistry , RWTH Aachen University , Forckenbeckstraße 50 , 52074 Aachen , Germany.
  • Ogieglo W; DWI - Leibniz Institute for Interactive Materials and Institute of Technical and Macromolecular Chemistry , RWTH Aachen University , Forckenbeckstraße 50 , 52074 Aachen , Germany.
  • Knoll W; Biosensor Technologies , AIT-Austrian Institute of Technology GmbH , Konrad-Lorenz-Straße 24 , 3430 Tulln an der Donau , Austria.
  • Rodriguez-Emmenegger C; DWI - Leibniz Institute for Interactive Materials and Institute of Technical and Macromolecular Chemistry , RWTH Aachen University , Forckenbeckstraße 50 , 52074 Aachen , Germany.
  • Dostálek J; Biosensor Technologies , AIT-Austrian Institute of Technology GmbH , Konrad-Lorenz-Straße 24 , 3430 Tulln an der Donau , Austria.
ACS Sens ; 4(8): 2109-2116, 2019 08 23.
Article em En | MEDLINE | ID: mdl-31364363
ABSTRACT
A compact optical biosensor for direct detection of thrombin in human blood plasma (HBP) is reported. This biosensor platform is based on wavelength spectroscopy of diffraction-coupled surface plasmons on a chip with a periodically corrugated gold film that carries an antifouling thin polymer layer consisting of poly[(N-(2-hydroxypropyl)methacrylamide)-co-(carboxybetaine methacrylamide)] (poly(HPMA-co-CBMAA)) brushes. This surface architecture provides superior resistance to nonspecific and irreversible adsorption of abundant compounds in the analyzed HBP samples in comparison to standard surface modifications. The carboxylate groups along the polymer brushes were exploited for the covalent immobilization of aptamer ligands. These ligands were selected to specifically capture the target thrombin analyte from the analyzed HBP sample in a way that does not activate the coagulatory process at the biosensor surface with poly(HPMA-co-CBMAA) brushes. Direct label-free analysis of thrombin in the medically relevant concentration range (1-20 nM) is demonstrated without the need for diluting the HBP samples or using additional steps for signal enhancement. The reported platform constitutes the first step toward a portable and sensitive point-of-care device for direct detection of thrombin in human blood.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombina / Técnicas Biossensoriais Limite: Humans Idioma: En Revista: ACS Sens Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Trombina / Técnicas Biossensoriais Limite: Humans Idioma: En Revista: ACS Sens Ano de publicação: 2019 Tipo de documento: Article