Your browser doesn't support javascript.
loading
Incidence of magnet mode in patients with implantable cardioverter defibrillators.
von Olshausen, Gesa; Schorr, Johanna; Grebmer, Christian; Weigand, Severin; Blazek, Patrick; Brkic, Amir; Friedrich, Lena; Semmler, Verena; Kolb, Christof; Lennerz, Carsten.
Afiliação
  • von Olshausen G; Klinik und Poliklinik für Innere Medizin I, Technische Universität München, Ismaninger Str. 22, 81675, Munich, Germany. gesa.olshausen@tum.de.
  • Schorr J; Klinik für Anästhesiologie, Klinikum rechts der Isar, Technische Universität München, Ismaninger Str. 22, 81675, Munich, Germany.
  • Grebmer C; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Weigand S; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Blazek P; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Brkic A; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Friedrich L; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Semmler V; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Kolb C; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
  • Lennerz C; Deutsches Herzzentrum München, Klinik für Herz- und Kreislauferkrankungen, Technische Universität München, Lazarettstr. 36, 80636, Munich, Germany.
J Interv Card Electrophysiol ; 56(3): 335-339, 2019 Dec.
Article em En | MEDLINE | ID: mdl-31617041
ABSTRACT

PURPOSE:

To determine the current incidence of antitachycardia therapy suspension due to magnet reversion in patients with implantable cardioverter defibrillators (ICDs).

METHODS:

From March 2012 to September 2018, all St. Jude Medical ICD interrogations performed at the Deutsches Herzzentrum München were screened for stored events of antitachycardia therapy suspension caused by static magnetic fields (flux densities ≥ 1 mT as defined in St. Jude Medical ICDs) affecting the giant magnetoresistance (GMR) sensor. Intentional ICD deactivation or inappropriate magnet application in the hospital was classified as non-environmental electromagnetic interference (EMI) whereas in the case of any other stored magnet reversion, the event was classified as presumably environmental EMI.

RESULTS:

Data from 201 St. Jude Medical ICD patients (mean age 62 ± 24 years, 165 males/36 females) were included in the analysis. ICD interrogations occurred at a mean follow-up of 25.1 months ± 15.6 months and summed up to a total of 464 patient-years of follow-up. A total number of 43 magnet reversion events were detected in 21 patients. Thirty-two events in 13 patients were presumably related to environmental EMI sources. Suspension of antitachycardia therapy only occurred in a temporary fashion. None of the patients experienced any ventricular tachyarrhythmia during ICD inhibition. The incidence of antitachycardia therapy suspension due to magnet reversion related to presumably environmental EMI sources was 6.9% per patient and year.

CONCLUSION:

Although none of the patients experienced any harmful event, antitachycardia therapy suspension due to magnet reversion is a common issue. Patients should be well-educated about potential EMI sources as well as trained in handling them.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Campos Eletromagnéticos Tipo de estudo: Etiology_studies / Incidence_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Card Electrophysiol Ano de publicação: 2019 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desfibriladores Implantáveis / Campos Eletromagnéticos Tipo de estudo: Etiology_studies / Incidence_studies / Risk_factors_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: J Interv Card Electrophysiol Ano de publicação: 2019 Tipo de documento: Article