The measurement of blood pressure by the linear method compared to the deflation methods differently modifies the pulse oximeter alarm frequency.
Kitamoto, Norihisa; Hitosugi, Takashi; Kawakubo, Yoshifumi; Hayashi, Keisuke; Yokoyama, Takeshi.
Afiliação
  • Kitamoto N; Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, Fukuoka.
  • Hitosugi T; Seirei hamamatsu General Hospital Department of Clinical Engineering, Hamamatsu, Japan.
  • Kawakubo Y; Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, Fukuoka.
  • Hayashi K; Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, Fukuoka.
  • Yokoyama T; Department of Dental Anesthesiology, Faculty of Dental Science, Kyushu University, Fukuoka.
Blood Press Monit ; 25(1): 13-17, 2020 Feb.
Article em En | MEDLINE | ID: mdl-31633521
ABSTRACT

BACKGROUND:

Noninvasive blood-pressure measurement device and pulse oximeter are important for patient monitoring. When these are placed on the same side, cuff inflation sometimes causes measurement failure by pulse oximeter.

OBJECTIVE:

The present study aimed to compare the pulse oximeter alarm frequency and pulse-wave disappearance duration between noninvasive blood-pressure measurement using the deflation method and that using the linear inflation method.

METHODS:

The study included 10 healthy subjects. The cuff for automatic sphygmomanometer was wrapped on one side of the upper arm and for pulse oximeter was attached to the thumb of the same side of upper limbs.

RESULTS:

The alarm frequency was 0 and 26% using the linear inflation and the deflation methods, respectively. Additionally, the pulse-wave disappearance duration was significantly longer using the deflation method than that using the linear inflation method (10.0 ± 1.5 vs 1.7 ± 0.8 s). With the linear inflation method, this duration was or less 3 s. In the deflation method, an excess pressure of 40 mmHg was used, which caused the alarm to turn on. Additionally, the heart rate was found to influence the alarm occurrence during measurement using the deflation method.

CONCLUSION:

Heart rate may influence alarm occurrence during blood-pressure measurement using the step deflation method. Using the linear inflation method, the risks of alarm occurrence and measurement failure are low, even when the pulse oximeter and blood-pressure measurement cuffs are installed on the same side, suggesting that this method is suitable for clinical use.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pressão Sanguínea / Determinação da Pressão Arterial / Oximetria / Esfigmomanômetros Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Blood Press Monit Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pressão Sanguínea / Determinação da Pressão Arterial / Oximetria / Esfigmomanômetros Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Blood Press Monit Ano de publicação: 2020 Tipo de documento: Article