Towards a generalised development of synthetic CT images and assessment of their dosimetric accuracy.
Handrack, Josefine; Bangert, Mark; Möhler, Christian; Bostel, Tilman; Greilich, Steffen.
Afiliação
  • Handrack J; Department of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Bangert M; Heidelberg Institute for Radiation Oncology, National Center for Radiation Research in Oncology, Heidelberg, Germany.
  • Möhler C; Department of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
  • Bostel T; Heidelberg Institute for Radiation Oncology, National Center for Radiation Research in Oncology, Heidelberg, Germany.
  • Greilich S; Department of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Acta Oncol ; 59(2): 180-187, 2020 Feb.
Article em En | MEDLINE | ID: mdl-31694437
ABSTRACT

Background:

The interest in generating "synthetic computed tomography (CT) images" from magnetic resonance (MR) images has been increasing over the past years due to advances in MR guidance for radiotherapy. A variety of methods for synthetic CT creation have been developed, from simple bulk density assignment to complex machine learning algorithms.Material and

methods:

In this study, we present a general method to determine simplistic synthetic CTs and evaluate them according to their dosimetric accuracy. It separates the requirements on the MR image and the associated calculation effort to generate a synthetic CT. To evaluate the significance of the dosimetric accuracy under realistic conditions, clinically common uncertainties including position shifts and Hounsfield lookup table (HLUT) errors were simulated. To illustrate our approach, we first translated CT images from a test set of six pelvic cancer patients to relative electron density (ED) via a clinical HLUT. For each patient, seven simplified ED images (simED) were generated at different levels of complexity, ranging from one to four tissue classes. Then, dose distributions optimised on the reference ED image and the simEDs were compared to each other in terms of gamma pass rates (2 mm/2% criteria) and dose volume metrics.

Results:

For our test set, best results were obtained for simEDs with four tissue classes representing fat, soft tissue, air, and bone. For this simED, gamma pass rates of 99.95% (range 99.72-100%) were achieved. The decrease in accuracy from ED simplification was smaller in this case than the influence of the uncertainty scenarios on the reference image, both for gamma pass rates and dose volume metrics.

Conclusions:

The presented workflow helps to determine the required complexity of synthetic CTs with respect to their dosimetric accuracy. The investigated cases showed potential simplifications, based on which the synthetic CT generation could be faster and more reproducible.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Imageamento por Ressonância Magnética Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Acta Oncol Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Imageamento por Ressonância Magnética Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Acta Oncol Ano de publicação: 2020 Tipo de documento: Article