Nano-etched fused-silica capillary used for on-line preconcentration and electrophoretic separation of bacteriophages from large blood sample volumes with off-line MALDI-TOF mass spectrometry identification.
Horká, Marie; Karásek, Pavel; Salplachta, Jirí; Ruzicka, Filip; Stveráková, Dana; Pantucek, Roman; Roth, Michal.
Afiliação
  • Horká M; Institute of Analytical Chemistry of the CAS, Veverí 97, 602 00, Brno, Czech Republic. horka@iach.cz.
  • Karásek P; Institute of Analytical Chemistry of the CAS, Veverí 97, 602 00, Brno, Czech Republic.
  • Salplachta J; Institute of Analytical Chemistry of the CAS, Veverí 97, 602 00, Brno, Czech Republic.
  • Ruzicka F; Department of Microbiology, Faculty of Medicine, Masaryk University, Kamenice 5, 625 00, Brno, Czech Republic.
  • Stveráková D; Department of Experimental Biology, Faculty of Science, Masaryk University, Kotlárská 2, 611 37, Brno, Czech Republic.
  • Pantucek R; Department of Experimental Biology, Faculty of Science, Masaryk University, Kotlárská 2, 611 37, Brno, Czech Republic.
  • Roth M; Institute of Analytical Chemistry of the CAS, Veverí 97, 602 00, Brno, Czech Republic.
Mikrochim Acta ; 187(3): 177, 2020 02 19.
Article em En | MEDLINE | ID: mdl-32076849
The properties of staphylococcal phages from the Siphoviridae, Podoviridae, and Myoviridae families were monitored using capillary electrophoretic methods on fused-silica capillaries with different morphology of surface roughness. Isoelectric points of the examined phages were determined by capillary isoelectric focusing in the original, smooth fused-silica capillary, and they ranged from 3.30 to 3.85. For capillary electrophoresis of phages, fused-silica capillaries with the "pock" and "cone" roughened surface types were prepared by etching a part of the capillary with supercritical water. The best resolution of the individual phages (to range from 3.2 to 4.6) was achieved with the "cone" surface-type fused-silica capillary. Direct application of phage K1/420 at the infection site, represented by human plasma or full blood spiked with Staphylococcus aureus, was on-line monitored by micellar electrokinetic chromatography. The phage particles were dynamically adhered onto the roughened surface of the capillary from 10 µL of the prepared sample at the optimized flow rate of 6.5 µL min-1. The limit of detection was determined to be 104 phage particles. The linearity of the calibration lines was characterized by the regression coefficient, R2 = 0.998. The relative standard deviation (RSD) of the peak area, calculated from ten independent measurements, was (±) 2%. After analysis, viability of the detected phages was verified by the modified "double-layer drop assay" method, and collected phage fractions were simultaneously off-line analyzed by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Graphical abstract.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bacteriófagos / Coleta de Amostras Sanguíneas / Dióxido de Silício / Espectrometria de Massas por Ionização e Dessorção a Laser Assistida por Matriz Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Mikrochim Acta Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Bacteriófagos / Coleta de Amostras Sanguíneas / Dióxido de Silício / Espectrometria de Massas por Ionização e Dessorção a Laser Assistida por Matriz Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Mikrochim Acta Ano de publicação: 2020 Tipo de documento: Article