Validation of the analytical performance of the NOVEOS™ System, a system which improves upon the third-generation in vitro allergy testing technology.
Bauersachs, Daniel; Potapova, Ekaterina; Renz, Harald; Benes, Stephanie Hagner; Matricardi, Paolo Maria; Skevaki, Chrysanthi.
Afiliação
  • Bauersachs D; Institute of Laboratory Medicine, Universities of Giessen and Marburg Lung Center (UGMLC), Philipps University Marburg, German Center for Lung Research (DZL), Marburg, Germany.
  • Potapova E; Department of Pediatric Pulmonology, Immunology and Intensive Care Medicine, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Renz H; Institute of Laboratory Medicine, Universities of Giessen and Marburg Lung Center (UGMLC), Philipps University Marburg, German Center for Lung Research (DZL), Marburg, Germany.
  • Benes SH; Institute of Laboratory Medicine, Universities of Giessen and Marburg Lung Center (UGMLC), Philipps University Marburg, German Center for Lung Research (DZL), Marburg, Germany.
  • Matricardi PM; Department of Pediatric Pulmonology, Immunology and Intensive Care Medicine, Charité - Universitätsmedizin Berlin, Berlin, Germany.
  • Skevaki C; Institute of Laboratory Medicine, Universities of Giessen and Marburg Lung Center (UGMLC), Philipps University Marburg, German Center for Lung Research (DZL), Marburg, Germany.
Clin Chem Lab Med ; 58(11): 1865-1874, 2020 10 25.
Article em En | MEDLINE | ID: mdl-32549134
ABSTRACT
Objectives Detection of allergen-specific immunoglobulin E (sIgE) is important for the diagnosis of allergy. IgE sensitization is commonly demonstrated in vivo by skin prick testing (SPT), or in vitro utilizing automated systems. Recently, HYCOR® Biomedical launched its new system for allergen sIgE testing called the NOVEOS™ Immunoanalyzer. This study aims to evaluate the analytical performance of the NOVEOS system in a bi-center study at Philipps-University Marburg (Site-1) and Charité Medical University Berlin (Site-2), respectively. Methods The analytical performance was evaluated based on the guidelines I/LA20-A3, EP5-A3, EP17-A2, EP6-A, EP7-A3, and EP9-A3 of the Clinical and Laboratory Standards Institute (CLSI). Results The conducted repeatability and within-laboratory precision tests provided acceptable performance with 3.0%-11.9% coefficient of variation across both sites. The limit of blank (LoB) and limit of detection (LoD) were <0.1 kU/L at both centers. A within-parameter linearity for all tested allergens was reported at both sites. Of note, no significant interference was observed for high levels of biotin, methylprednisolone, diphenhydramine, omalizumab, or ranitidine. Method comparison between the NOVEOS calibration and the latest World Health Organization (WHO) reference standard showed good agreement at both sites. Conclusions The results from the analytical performance of the NOVEOS allergen sIgE assay and instrument testing at both sites were comparable. Overall, a good precision and linearity as well as a detection limit <0.1 kU/L were observed, with minimal impact of common interfering substances on patient recoveries. The NOVEOS is calibrated to the latest WHO reference standard and adds benefits like a small sample size and para-magnetic microparticles that improve upon third-generation allergen sIgE assays' design and performance.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 / 2_ODS3 Base de dados: MEDLINE Assunto principal: Imunoglobulina G / Hipersensibilidade Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline Limite: Humans Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 1_ASSA2030 / 2_ODS3 Base de dados: MEDLINE Assunto principal: Imunoglobulina G / Hipersensibilidade Tipo de estudo: Clinical_trials / Diagnostic_studies / Guideline Limite: Humans Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2020 Tipo de documento: Article