Comparison of the reliability of Gram-negative and Gram-positive flags of the Sysmex UF-5000 with manual Gram stain and urine culture results.
Enko, Dietmar; Stelzer, Ingeborg; Böckl, Michael; Schnedl, Wolfgang J; Meinitzer, Andreas; Herrmann, Markus; Tötsch, Martin; Gehrer, Michael.
Afiliação
  • Enko D; Clinical Institute of Medical and Chemical Laboratory Diagnostics, Medical University of Graz, Graz, Austria.
  • Stelzer I; Institute of Clinical Chemistry and Laboratory Medicine, General Hospital Hochsteiermark, Leoben, Austria.
  • Böckl M; Institute of Clinical Chemistry and Laboratory Medicine, General Hospital Hochsteiermark, Leoben, Austria.
  • Schnedl WJ; Institute of Clinical Chemistry and Laboratory Medicine, General Hospital Hochsteiermark, Leoben, Austria.
  • Meinitzer A; Practice for General Internal Medicince, Bruck/Mur, Austria.
  • Herrmann M; Clinical Institute of Medical and Chemical Laboratory Diagnostics, Medical University of Graz, Graz, Austria.
  • Tötsch M; Clinical Institute of Medical and Chemical Laboratory Diagnostics, Medical University of Graz, Graz, Austria.
  • Gehrer M; Institute of Pathology, General Hospital Hochsteiermark, Leoben, Austria.
Clin Chem Lab Med ; 59(3): 619-624, 2021 02 23.
Article em En | MEDLINE | ID: mdl-33068381
ABSTRACT

OBJECTIVES:

Recently, the fully automated flow cytometry-based UF-5000 (Sysmex Corboration, Kobe, Japan) urine sediment analyzer was developed providing bacteria (BACT) info flags for more accurate bacterial discrimination of urinary tract infections (UTIs). This study aimed to compare the reliability of the UF-5000 BACT-info flags with manual Gram stain and urine culture as the gold standard method.

METHODS:

A total of 344 urine samples were analyzed on the UF-5000 and compared with manual microscopic Gram stain and urine cultures. Agreement was assessed by Cohen's kappa (κ) analysis. The Youden index was used to determine the optimal BACT and white blood cell (WBC) cut-off points for discriminating positive and negative urine cultures.

RESULTS:

Overall 98/344 (28.5%) samples were urine culture positive at a cut-off of ≥105 CFU/mL. "Gram-negative?" UF-5000 BACT-Info flags showed a better concordance of 25/40 (62.5%) with urine culture compared to Gram stain with 30/50 (60%). The results for UF-5000 discrimination of Gram-positive and Gram-negative microorganisms demonstrated a substantial (κ = 0.78) and fair (κ = 0.40) agreement with urine culture. Optimal cut-off points detecting positive urine cultures were 135 BACT/µL (sensitivity [SE] 92.1%, specificity [SP] 85.4%, positive predictive value [PPV] 71%, negative predictive value [NPV] 96%) and 23 WBC/µL (SE 73.5%, SP 84.1%, PPV 65%, NPV 89%).

CONCLUSIONS:

The UF-5000 analyzer (Sysmex) is a reliable diagnostic tool for UTI screening. The displayed BACT-Info flags allow a quick diagnostic orientation for the clinician. However, the authors suggest verifying the automated Gram categories with urine culture.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções Urinárias Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções Urinárias Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies Limite: Humans Idioma: En Revista: Clin Chem Lab Med Ano de publicação: 2021 Tipo de documento: Article