Design and construction of a simplified, gas-driven, pressure-controlled emergency ventilator.
Szlosarek, R; Teichert, R; Wetzel, A; Fichtner, A; Reuter, F; Kröger, M.
Afiliação
  • Szlosarek R; Technische Universität Bergakademie Freiberg, Institute for Machine Elements, Design and Manufacturing, Freiberg, Germany.
  • Teichert R; Technische Universität Bergakademie Freiberg, Institute for Machine Elements, Design and Manufacturing, Freiberg, Germany.
  • Wetzel A; Technische Universität Bergakademie Freiberg, Institute for Machine Elements, Design and Manufacturing, Freiberg, Germany.
  • Fichtner A; Kreiskrankenhaus Freiberg gGmbH, Emergency Department, Freiberg, Germany.
  • Reuter F; Technische Universität Bergakademie Freiberg, Research and Training Mine Reiche Zeche, Freiberg, Germany.
  • Kröger M; Technische Universität Bergakademie Freiberg, Institute for Machine Elements, Design and Manufacturing, Freiberg, Germany.
Afr J Emerg Med ; 11(1): 175-181, 2021 Mar.
Article em En | MEDLINE | ID: mdl-33194539
ABSTRACT

INTRODUCTION:

Due to the COVID-19 crisis or any other mass casualty situation it might be necessary to give artificial ventilation to many affected patients. Contrarily, the worldwide availability of emergency ventilators is still a shortage, especially in developing countries.

METHODS:

Modes of artificial ventilation were compared and the most safe, easy to use, and lung protecting principle was optimized to fit all requirements of both emergency ventilation and cost-effective mass production.

RESULTS:

The presented research results describe a simplified device for a pressure-controlled ventilation which works without electricity according to a known principle. Just pressurized gas and a patient connection is required. The device enables the control of basic ventilator parameters such as peak inspiratory pressure, positive end-expiratory pressure and the ventilation frequency. Further, the device is semiadaptive to the patient's lung stiffness and automatically maintains minute volume through frequency adjustment. The machine can be manufactured by turning, milling and drilling and needs purchased components with costs less than 100 USD. A sterilization and thus a reuse is possible.

DISCUSSION:

The presented development does not describe a ready-to-purchase ventilator, it rather outlines a refined working principle for emergency ventilation and its easiest methods of production with a minimum of requirements. The presented research aims on providing an open-source guideline for production of an emergency ventilator using worldwide available methods and thus should inspire local researchers to do a reverse engineering and eventually to put it into operation following country-specific regulations. For long-term ventilation exceeding emergency purposes, a monitoring of alarms for disconnection and violation of desired ventilator parameters should be established. The ventilator is limited to a fixed ratio between PIP and PEEP. Moreover, the ventilation frequency depends on two parameters, which needs some training. Nevertheless, the ventilator provides basic features to enable an emergency ventilation with minimal prerequisites.