Automatic detection of valve events by epicardial accelerometer allows estimation of the left ventricular pressure trace and pressure-displacement loop area.
Wajdan, Ali; Krogh, Magnus Reinsfelt; Villegas-Martinez, Manuel; Halvorsen, Per Steinar; Elle, Ole Jakob; Remme, Espen Wattenberg.
Afiliação
  • Wajdan A; The Intervention Centre, Oslo University Hospital, Rikshospitalet, Oslo, Norway.
  • Krogh MR; Department of Informatics, University of Oslo, Oslo, Norway.
  • Villegas-Martinez M; The Intervention Centre, Oslo University Hospital, Rikshospitalet, Oslo, Norway.
  • Halvorsen PS; Department of Informatics, University of Oslo, Oslo, Norway.
  • Elle OJ; The Intervention Centre, Oslo University Hospital, Rikshospitalet, Oslo, Norway.
  • Remme EW; Institute of Clinical Medicine, Faculty of Medicine, University of Oslo, Oslo, Norway.
Sci Rep ; 10(1): 20088, 2020 11 18.
Article em En | MEDLINE | ID: mdl-33208784
ABSTRACT
Measurements of the left ventricular (LV) pressure trace are rarely performed despite high clinical interest. We estimated the LV pressure trace for an individual heart by scaling the isovolumic, ejection and filling phases of a normalized, averaged LV pressure trace to the time-points of opening and closing of the aortic and mitral valves detected in the individual heart. We developed a signal processing algorithm that automatically detected the time-points of these valve events from the motion signal of a miniaturized accelerometer attached to the heart surface. Furthermore, the pressure trace was used in combination with measured displacement from the accelerometer to calculate the pressure-displacement loop area. The method was tested on data from 34 animals during different interventions. The accuracy of the accelerometer-detected valve events was very good with a median difference of 2 ms compared to valve events defined from hemodynamic reference recordings acquired simultaneously with the accelerometer. The average correlation coefficient between the estimated and measured LV pressure traces was r = 0.98. Finally, the LV pressure-displacement loop areas calculated using the estimated and measured pressure traces showed very good correlation (r = 0.98). Hence, the pressure-displacement loop area can be assessed solely from accelerometer recordings with very good accuracy.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Função Ventricular Esquerda / Pressão Ventricular / Acelerometria / Valvas Cardíacas / Hemodinâmica Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: Sci Rep Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Função Ventricular Esquerda / Pressão Ventricular / Acelerometria / Valvas Cardíacas / Hemodinâmica Tipo de estudo: Diagnostic_studies Limite: Animals Idioma: En Revista: Sci Rep Ano de publicação: 2020 Tipo de documento: Article