Your browser doesn't support javascript.
loading
Development, characterization, and testing of a personal passive sampler for measuring inhalation exposure to gaseous elemental mercury.
Snow, Melanie A; Feigis, Michelle; Lei, Ying Duan; Mitchell, Carl P J; Wania, Frank.
Afiliação
  • Snow MA; Department of Physical and Environmental Sciences, University of Toronto Scarborough, 1265 Military Trail, Toronto, Ontario M1C 1A4, Canada.
  • Feigis M; Department of Physical and Environmental Sciences, University of Toronto Scarborough, 1265 Military Trail, Toronto, Ontario M1C 1A4, Canada.
  • Lei YD; Department of Physical and Environmental Sciences, University of Toronto Scarborough, 1265 Military Trail, Toronto, Ontario M1C 1A4, Canada.
  • Mitchell CPJ; Department of Physical and Environmental Sciences, University of Toronto Scarborough, 1265 Military Trail, Toronto, Ontario M1C 1A4, Canada.
  • Wania F; Department of Physical and Environmental Sciences, University of Toronto Scarborough, 1265 Military Trail, Toronto, Ontario M1C 1A4, Canada. Electronic address: frank.wania@utoronto.ca.
Environ Int ; 146: 106264, 2021 01.
Article em En | MEDLINE | ID: mdl-33227582
ABSTRACT
Inhalation of gaseous elemental mercury (GEM) is an occupational exposure concern for workers handling elemental mercury or mercury-containing waste. GEM is also often present near historically mercury-contaminated sites, potentially resulting in low-level, chronic exposure of the wider population. Here we introduce a passive sampler for personal GEM monitoring which combines a radial porous diffusive barrier with an activated carbon sorbent. A total mercury analyzer is used to quantify GEM sorbed to the carbon by thermal decomposition, amalgamation, and atomic absorption spectroscopy. A sampling rate of 0.070 m3/day was determined by calibrating the sampler at low and high concentrations. Deployments lasting 8 h result in limits of quantification well below 200 ng/m3. The sampler has a measurement range of at least four orders of magnitude. Derived air concentrations were not statistically significantly different from those obtained by active air sampling but were more precise than those obtained using a personal pump. If properly stored, the sampler maintains low blank levels in high GEM environments. Affordability, sturdiness, simplicity, and the wide availability of total mercury analyzers make this sampler highly suited for monitoring GEM inhalation exposure, including in developing countries.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 2_ODS3 Base de dados: MEDLINE Assunto principal: Exposição Ocupacional / Poluentes Atmosféricos / Mercúrio Limite: Humans Idioma: En Revista: Environ Int Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Contexto em Saúde: 2_ODS3 Base de dados: MEDLINE Assunto principal: Exposição Ocupacional / Poluentes Atmosféricos / Mercúrio Limite: Humans Idioma: En Revista: Environ Int Ano de publicação: 2021 Tipo de documento: Article